SigmaPaceTM 1000: Medtronic Temporary External Pacemaker Model 5388 Checkout Procedure
1. PRELIMINARY SET UP
a. Connect the pacemaker output to the analyzer’s ATRIAL and VENTRICULAR 4 mm interface jacks using the supplied (red and black) test leads.
b. Turn on power on both the pacemaker and the analyzer.
From the main (power-up) menu on the analyzer:
Press F2: INVasive pacemakers
Press F1: NEXT (Tests pacemaker using the default 500 Ω test load.)
Remaining menu navigation directions are given from this Ext-Invasive
Pacer SELECT TEST screen.
Note:
Use the F1: HOLD/RELEASE key feature to freeze test results on the analyzer’s screen.
Use the analyzer’s ESCape key to EXIT out of specific test modes.
2. RATE TEST
a. ARTRIAL RATE
(Use SigmaPace 1000 F1: PULSE OUTPUT, F3: DUAL A+V Screen)
Operational Mode: AAI
Atrial Channel: ON and Ventricular Channel: OFF by pressing the key button on the 5388 and setting the V output to zero.
Atrial Channel Output @ 10 mA
Press the key button on the 5388 and check the rate settings.
Rate Setting | Specified Tolerance (PPM) | Specified Tolerance (mS) | SigmaPace 1000 Reading | Status Pass / Fail? |
---|---|---|---|---|
30 PPM | ± 5 % (28.5 – 31.5) | 1900 – 2100 | PPM | |
80 PPM | ± 5 % (76 – 84) | 712 – 788 | PPM | |
120 PPM | ± 5 % (114 – 126) | 475 – 525 | PPM | |
200 PPM | ± 5 % (190 – 210) | 285 – 315 | PPM |
b. VENTRICULAR RATE
(Use SigmaPace 1000 F1: PULSE OUTPUT, F3: DUAL A+V Screen)
Operational Mode: VVI
Atrial Channel: OFF by pressing the key button on the 5388 and setting the
A output to zero. and Ventricular Channel: ON
Ventricular Channel Output @ 10 mA
Press the key button on the 5388 and check the rate settings.
Rate Setting | Specified Tolerance (PPM) | Specified Tolerance (mS) | SigmaPace 1000 Reading | Status Pass / Fail? |
---|---|---|---|---|
30 PPM | ± 5 % (28.5 – 31.5) | 1900 – 2100 | PPM | |
80 PPM | ± 5 % (76 – 84) | 712 – 788 | PPM | |
120 PPM | ± 5 % (114 – 126) | 475 – 525 | PPM | |
200 PPM | ± 5 % (190 – 210) | 285 – 315 | PPM |
c. RAPID ATRIAL RATE (RAP)
(Use SigmaPace 1000 F1: PULSE OUTPUT, F3: DUAL A+V Screen)
Operational Mode: AOO (Access RAP feature via the Medtronic 5388 Menu #3).
Atrial Channel: ON and Ventricular Channel: OFF
Atrial Channel Output @ 10 mA
Rate Setting | Specified Tolerance (PPM) | Specified Tolerance (mS) | SigmaPace 1000 Reading | Status Pass / Fail? |
---|---|---|---|---|
350 PPM | ± 5 % (285 – 367.5) | 163 – 180 | PPM |
3. PULSE WIDTH
a. ATRIAL PULSE WIDTH
(Use SigmaPace 1000 F1: PULSE OUTPUT, F3: DUAL A+V Screen)
Operational Mode: AAI
Atrial Channel: ON and Ventricular Channel: OFF
Atrial Channel Output @ 10 mA
Rate Setting | Pulse Width | Specified Tolerance (mS) | SigmaPace 1000 Reading | Status Pass / Fail? |
---|---|---|---|---|
80 PPM | 1.0 mS | ± 5 % (0.95 – 1.05) | mS |
b. VENTRICULAR PULSE WIDTH
(Use SigmaPace 1000 F1: PULSE OUTPUT, F3: DUAL A+V Screen)
Operational Mode: VVI
Atrial Channel: OFF and Ventricular Channel: ON
Ventricular Channel Output @ 10 mA
Rate Setting | Pulse Width | Specified Tolerance (mS) | SigmaPace 1000 Reading | Status Pass / Fail? |
---|---|---|---|---|
80 PPM | 1.5 mS | ± 5 % (1.425 – 1.575) | mS |
4. AV INTERVAL
(Use SigmaPace 1000 F1: PULSE OUTPUT, F3: DUAL A+V Screen)
Operational Mode: DDD (Access this feature via the Medtronic 5388 Menu #1 or use it’s power up default output settings.)
Atrial Channel: ON and Ventricular Channel: ON
Atrial Channel Output @ 10 mA
Ventricular Channel Output @ 10 mA
Rate Setting | AV Interval Setting | Specified Tolerance (mS) | SigmaPace 1000 Reading | Status Pass / Fail? |
---|---|---|---|---|
80 PPM | 20 mS | ± 5 mS (15 – 25) | mS | |
80 PPM | 150 mS | ± 5 % (142 – 158) | mS | |
80 PPM | 300 mS* | ± 5 % (285 – 315) | mS |
*To set the Medtronic 5388 AV interval @ 300 mS, first decrease the PVARP to minimum (150 mS) on MENU #2 and then go to Menu #1 to access the AV interval function.