Why am I not able to get files from ProSim 8 using Ansur ProSim 8 Mini Plug-In?

Ansur ProSim 8 mini Plug-In requires administrator privileges to run properly. If you have not logged on to your PC as administrator, press the SHIFT key on the keyboard while right-clicking on the Ansur shortcut or ProSim 8 mini Plug-In shortcut on the desktop or Start Menu, then select "run as" administrator. You will now be able to run Ansur ProSim Mini Plug-In without any errors.

When to use autosequence container and step in Ansur ProSim Plug-In?

a) If you want to execute tests sequentially one by one and enter the results individually and navigate to next test element manually then, drag and drop the test elements into the template without autosequence container
b) If you want to execute tests sequentially one by one and enter the results individually and navigate to next test element automatically then, drag and drop the test elements into the template under the autosequence container

What is the difference between execution of tests in serial and parallel in Ansur ProSim <br>Plug-In?

Execution of tests in serial allows the user to simulate waveforms individually, and ProSim device resets to LOCAL mode at completion of each test
b) Execution of the tests in parallel allows user to simulate tests simultaneously, and ProSim device resets to LOCAL mode only at the completion of each Step

How to create autosequence from Ansur similar to the autosequence present in the <br>ProSim simulator?

To create autosequence similar to the autosequence present in the ProSim
simulator follow the steps below
a) Drag and drop Autosequence container to the template
b) Drag and drop one more autosequence container under it. This becomes the step which simulates the test elements in parallel
c) Drag and drop test elements under the step
d) Save it as custom templates in the Ansur test library or to any other location in PC

Can I use all the test elements of Ansur ProSim Plug-In inside a step?

a) Step can have following test elements under it
i. Any one test element from the ECG Simulation Test group
ii. Respiration Simulation Test
iii. Temperature Simulation Test
iv. Any one of IBP Dynamic Simulation or IBP Static pressure test from the IBP Simulation Test
group
&nnbsp;v. Only NIBP Dynamic Simulation test from the NIBP Test group
vi. SpO2 test
b) If you include test elements other than the above, Ansur displays a message box and discards those test elements