Defibrillator Analyzers and AED Testers

Will the Dale 900/ Impulse 3000 test biphasic defibs?

Yes, all Dale 900 defib analyzers will test biphasic defibs. The Impulse 3000 will test biphasic defibs depending on the firmware version. The firmware version is briefly displayed at power up. Firmware versions earlier than version 1.08 were not tested but are considered to be incompatible. Version 1.10 is also not compatible. These versions will generally give you inconsistent and usually low joule output readings when you fire your biphasic defib. Versions 1.11 higher will be compatible with biphasic defibs.

When using the defib module on the medtester 5000C and I discharge the defibrillator, the joule values are displayed on the Impulse 3000 or Dale 900 screen but not on the medtester screen. What is wrong?

There is a MEDT selection on the menus of the Impulse 3000 and Dale 900. Make sure that you have pushed this button, that the baud rate is set to 2400 and that you are using the proper cable. The correct cable to use with the Impulse 3000 and Dale 900 is Fluke part number 2199346. The older part number was 3010-0300.

What is the purpose of the TEST button on the Dale 900/Impulse 3000 ENERGY menu?

The ENERGY menu of the Dale 900/Impulse 3000 has a TEST button that is used as a self-test or basic circuitry test. With the low range selected, the Joules reading on the lower left line of the Dale 900/Impulse 3000 display should read ~4.0 J. With the high range selected, the Joules reading should read 100.0 J +/- 4.0 J. While this is not an absolute calibration test, if your Dale 900/Impulse 3000 is starting to drift away from these values you should send your Dale 900/Impulse 3000 in to the Fluke service department for calibration.

Regarding the Defibrillator Test Plug-In, why does my Ansur IMPULSE 4000 Plug-In fail to communicate/setup my IMPULSE 4000 when I try to run a test template to test my defibrillator/pacer?

The IMPULSE 4000 firmware requires the display screen to be at the home-screen in order for communications to be established. On the IMPULSE 4000, press the ESC button (middle right on control panel) until the display shows IMPULSE 4000 MAIN MENU PAGE 1: Then run the test template. If you have other problems it is likely a loose or misconnected RS232 serial communications cable or an incorrect COM port selection on your computer. This generally applies to all Ansur-compatible test devices. Be sure that the display menu is at the home position on the compatible test device.