Skip to main content
Defibrillator analyzer representative icon
Defibrillator Analyzers and AED Testers

What is the purpose of the TEST button on the Impulse 4000 defib menu?

The defib menu of the Impulse 4000 has a TEST button that is used as a self-test or basic circuitry test. With the low range selected, the E or energy reading on the top right line of the Impulse 4000 display should read 4.0 J +/- 0.2 J. With the high range selected, the E or energy reading should read 100.0 J +/- 5.0 J. Please this is not an absolute calibration test. If your Impulse 4000 is starting to drift away from these values, you should send your Impulse 4000 to the Fluke service department for calibration.

What is the part number for the paddle adapters for testing AED or hands free defibrillator output on Agilent/Philips defibrillators with the flat rectangular connector?

The paddle electrode adapters are very useful for testing AED or hands free defib testing. We do not recommend sticking defib electrode patches directly to the pads on the Impulse 4000. The part number for the Agilent/Philips paddle electrode adapter for the Impulse 4000 is 2201017.

What is the part number for the paddle adapters for testing AED or hands free defib output on Zoll Medical PD series, M series and AED Plus defibrillators with the two pin white connection?

The paddle electrode adapters are very useful for testing AED or hands free defib testing. We do not recommend sticking defib electrode patches directly to the pads on the Impulse 4000. The part number for the Zoll Medical PD series, M series and AED Plus paddle electrode adapter is 2199758.

What is the part number for the paddle adapters for testing AED or hands free defib output for Medtronics Physio Control defibrillators with the "Quik Combo" connection?

The paddle electrode adapters are very useful for testing AED or hands free defib testing. We do not recommend sticking defib electrode patches directly to the pads on the Impulse 4000. The part number for the Medtronics Physio Controls paddle electrode adapter is 2392249.

Should I discharge the defibrillator while performing pacer tests with the pacer module inserted into the Impulse 4000?

No, there is no need to discharge the defibrillator when testing the pacer on a defibrillator. There is also a chance that you could damage the Impulse 4000, the pacer module or both if you discharge the defibrillator while the pacer module is inserted into the Impulse 4000.

I have no problem connecting the 5000C to the Impulse 4000 during defib tests using the 5000C defib module, but the pacer module gives me a message that says "No Impulse 4000 Detected". Why won't the pacer module connect to the Medtester 5000C?

Make sure that you are using COMM 2 on the medtester. The correct cable to use is Fluke part number 2200252. The older part number was 3010-0467. The baud rates on both COMM 2 of the Medtester and the Impulse 4000 need to be set at 2400 baud for the connection to be successful. When you use the defib module on the 5000C the baud rate should automatically change from any baud rate to 2400 baud by itself. Therefore whatever the baud rate was set at for COMM 2 you will still get a connection. This is not true when using the pacer module on the medtester.