PS410
Patient Simulator
High-performance arrhythmia cardiac simulator device for patient monitor testing.

Key Features
- Compact, lightweight, pocket-sized
- Easy top-panel control
- 12-lead Electrocardiogram
- Adult and pediatric normal-sinus rhythms
- 35 arrhythmia selections
- Electrocardiogram performance waveforms
- ST segment levels
- Pacemaker simulation
- Electrocardiogram artifact
- Battery operated

High-performance arrhythmia cardiac simulator device for patient monitor testing.
This handheld device simulates a full range of cardiac rhythms and a wide variety of electrocardiogram conditions. It includes pacemaker simulation, 35 arrhythmia selections, and adult and pediatric normal-sinus rhythms.
The PS410 is a compact, high-performance arrhythmia cardiac simulator device for patient monitor testing. Small enough to fit in a pocket, the handy PS410 weighs less than a pound and is easy to operate. Technicians simply connect the PS410 to the device under test and use the PS410 keypad to enter the code presets. The PS410 then transmits the selected preset simulations to the device under test.
Documentation
Manuals
PS410 Security Procedures
PS410 Simulator Datasheet
PS410 User Manual
Available in English, Chinese, French, German, Italian, Japanese, Spanish
PS410 User's Manual Supplement
Datasheet
PS410 Simulator Datasheet
Certifications
PS410 CE Declaration of Conformity
Models

PS410 Arrhythmia Simulation Device
PN: 2631276
Frequently asked questions
Ventricular Fibrillation (VFIB) is a medical emergency condition in which there is uncoordinated contraction of the cardiac muscle of the ventricles in the heart, making them tremble rather than contract properly. Since the VFIB waveform continuously changes in shape, magnitude and direction, different patient monitors may use slightly different algorithms to detect VFIB signal.
When running the PS410/420's electronically-generated VFIB selection on either the SpaceLabs Healthcare UltraView 1500 or the Datascope Corporation Passport 2 patient monitors, the signal is not always recognized as VFIB. Randomly, short waveform segments may or may not be detected after 30 seconds to 4 minutes. Different algorithms are utilized by the range of available brands and models of patient monitoring equipment.
In this situation, we recommend customers use the FBC MPS450 Multi-Parameter Patient Simulator's VFIB selection with a broader frequency spectrum simulation signal recognizable by the SpaceLabs Healthcare UltraView 1500, Datascope Corporation Passport 2 and other brands/models of patient monitoring equipment used throughout the healthcare delivery system.