Workflow Automation Software
What is the difference between the Excel Add-In and Ansur Test Automation Software?
Both the Excel add-in and Ansur Test Automation software automatically capture test data and facilitate data management and reporting. The Excel add-in collects measured values and populates a customizable Microsoft Excel spreadsheet template; A MQSA-specific template is included for Mammo applications.
How can the firmware and/or software be updated by the end-user? Is there a fee?
Any firmware or software upgrades that do not affect calibration of the TNT 12000 are available on the Fluke Biomedical website under Support: Software Downloads.
How accurate is the waveform in the Excel Add-in or Ansur Test Guide (Ansur Plug-in)?
The waveform is accurate enough to assess whether there are problems in the X-ray tube or generator not disclosed by the numeric measurements. To obtain more accurate waveforms or to perform calibrations using the waveforms the 35080M199 X-Ray Scopemeter is recommended.