Defibrillator Analyzers and AED Testers

Using the defib module on the medtester 5000C when I fire the defibrillator the joule values are displayed on the Impulse 4000 screen but not on the medtester screen. What is wrong?

On Impulse 4000 Main Menu Page 2 there is a MEDT button that needs to be pushed to transfer data back to the medtester. Make sure that you have entered the baud rate at 2400 and that you are using the proper cable. The correct cable to use is Fluke part number 2200252. The older part number was 3010-0467.