Skip to main content
Biomedical Test Equipment family

Miscellaneous

I have an American Edwards disposable single patient use transducer and a cable connecting that transducer to the IBP channel of the patient monitor. How do I connect this kind of transducer and its cable to test them?

The focus of IBP testing is on the patient monitor IBP functionality from the point of connection to the patient monitor to the displayed and strip chart printed output. The accuracy of the measurement processing is part of this performance testing and safety testing. IBP and ICP accuracy are determined by a series of static pressure level comparisons.