VT Plus HF App Note: Ventilator Compensation Mode Settings

Ventilator Compensation Mode Settings

Manufacturer Model Correction Mode Comment
Viasys/Bird Avion BTPS  
Viasys/Bird Gold BTPS  
Viasys/Bird T-Bird Legacy BTPS  
BMD IC2A ATP  
BMD MVP10 ATP  
BMD Crossvent 4 ATP  
Breas PV-403, 501 ATP Not available from manufacturer, assumed
Breas IMP-2 ATP Not available from manufacturer, assumed
Draeger Evita 2 and 4 BTPS (or NTPD) Uses BTPS in normal mode but both in Service mode, NTPD for high pressure barometer (use STPD21 in this case)
Draeger Babylog 8000 0 BTPS (or NTPD) Uses BTPS in normal mode but both in Service mode, NTPD for high pressure barometer (use STPD21 in this case)
Hamilton Galileo 2 ATP  
Hamilton Veolar 3 ATP  
Infrasonics Adult Star STPD 21°  
Newport E100i & E100m ATP  
Newport Breeze E150 ATP Use BTPS when connecting after the water trap and heat up
Puritan Bennett 7200 BTPS  
Puritan Bennett 760 BTPS  
Puritan Bennett 840 BTPS  
Puritan Bennett MA-I BTPS  
Pulmonetics LTV Series BTPS  
Maquet (formerly Siemens) Servo i STPD 0°  
Maquet (formerly Siemens) 300C STPD 21°  
Maquet (formerly Siemens) 900C ATP  
       

Setting of BTPS (on the VT PLUS HF) will give higher readings on flow (and tidal volumes) than ATP, STP21 and STP0 (in order).

Definitions:

STPD: Standard Temperature and Pressure, Dry (0ºC or 21ºC)

BTPS: Body Temperature and Pressure, Saturated (think of warm, moist exhaled air)

ATP: Ambient Temperature and Pressure

NTPD: Normal Temperature and Pressure, Dry (see STPD at 21ºC)

 