Electrical safety test on an MRI in 10 easy steps
How to run an electrical safety test on an MRI in 10 easy steps with the ESA615 using the Point to Point function
Step 1
Plug the ESA615 into an electrical outlet
Step 2
Connect the alligator clips to both lead cords
Step 3
Plug the lead cord accessory (50’ or 75’) into the red input plug on the front of the ESA615
Step 4
Plug the lead cord with the black input jack into the black input plug on the front of the ESA615
Step 5
Put the alligator clips together and zero out the leads on the ESA615 by choosing Point to Point resistance.
Step 6
Insert the ground adapter into the electrical outlet for the ground reference point
Step 7
Connect the black alligator clip/lead cord to the ground adapter
Step 8
Run the long lead cord with the red alligator clip into the MRI room and clamp onto any exposed metal (ground point) on the MRI machine
Step 9
Run Point to Point to test function (volts, ohms, micro amps)
Step 10
Record results
You will need the following to perform this test:
Models/descriptions
4132046 ESA615 Electrical Safety Analyzer
(US/120V - alligator clamps included)
2392639 Ground adapter
4316223 50’ Lead cord
4794343 75’ Lead cord