RF303RS
Electrosurgery Analyzer

Electrosurgery safety and performance testing made easy
Image
Features

Signal Averaging Mode (SAMTM)

RF303RS Electrosurgical Analyzer

You can perform eletrosurgical safety maintenance checks on most units on the market today using the RF303RS. Compact and portable, the device is so simple to use that technicians can become proficient with the RF303RS within minutes. This electrosurgical analyzer measures ESU output and high-frequency leakage, allows verification tests on the return electrode contact quality monitors, and has an oscilloscope output for waveform viewing. Instantaneous output or selectable sample times provide extra versatility. The instantaneous mode is sufficient for most units, but if output readings are variable and require stabilizing, the signal-averaging mode allows users to manually select two additional, slower sampling times to produce an accurate average reading.

Documentation

Manuals

rf303 Operator's Manual

rf303 Operator's Manual Supplement

RF303 RS electrosurgery analyzer

Datasheet

RF303 RS electrosurgery analyzer

Models

What comes in the box:

  • Users manual (3031000)
  • Accessory kit (safety leads) (one each – active safety lead (TEST LEAD)
  • ESU dispersive safety lead (3010-0576)
  • ESU RECQM safety lead (3010-0575)
  • ESU case safety lead(TEST LEAD)
  • two each – ESU jumper safety lead (3010-0578)
  • Active safety clip yellow (2720-0005)
  • Case safety clip green (2710-0006)
  • Fuses (2) 5X20 F3.15A 250V CE (1005-0194)
  • Ground pin adapter (9503-0004)
  • Detachable power cord (country specific)
View all models

Frequently asked questions

Yes, test results can be printed by a serial printer (Such as the DPU-414, P/N 2248899) which is connected to the RF303RS serial port.
  • Put your RF303RS into "simplex" mode. To do this:
    • With the RF303RSturned off, hold down the "+" and "-" softkeys (used to select load resistance)
    • With the "+" and "-" soft keys depressed, turn on the RF303RS. Wait a moment, then release the soft keys
  • Your RF303RS is now in simplex mode. Using a straight-through serial cable, connect your printer to the RF303RS serial port. The RF303RS serial port is configured as follows: Baud rate is fixed at 2400. No parity, 1 stop bit. Configure your printer to the same settings the RF303RS uses, as shown above.
    • Turn on your printer. Put it "on line".
    • Take a power measurement reading. While the reading is being taken, push the "mode" button.
    • Data will be sent to your printer showing power in watts and current, as well as load resistance. Carriage Return and Line-feed signals will be sent to your printer.
    • Every time you push the "mode" button, one line of data will be sent to your printer.