PS320 Fetal Simulator
The PS320 Fetal Simulator simulates fetal and maternal ECG as well as uterine activity to test and troubleshoot fetal electronic monitors and to train clinical staff.
The unit is battery operated and small enough to fit in a pocket so mobile technicians and clinical instructors can take it anywhere.
The PS320 simulates several fetal parameters including twins, and it simulates a wide range of clinical scenarios for training labor-and-delivery staff in how to recognize normal and abnormal responses. An optional mechanical heart creates fetal heart sounds for testing fetal monitors' ultrasound cables and transducers.
PS320 offers an easy user interface providing a 2 x 16-character LCD display with
adjustable contrast. The unit operates on a 9-volt battery with low-battery monitoring or functions with the supplied battery eliminator.
|
|
Key Features
- Mechanical heart for ultrasound simulation
- TOCO simulation (External or IUP)
- Ultrasound simulation (including twins)
- Maternal ECG simulation
- Fetal ECG (tracks ultrasound #1)
- Internal (DECG) and external fetal ECG
- Uterine-activity selections
- Fetal beat-to-beat variability
- Periodic and non-periodic fetal ECG changes
- Arrhythmia selections
- Compact, lightweight, pocket-size plastic housing
- Battery operated with status indications
- Special kits available with all required accessories and cables to test fetal monitors for specified manufacturers
|
|
|
Specifications
| Fetal ECG |
| Static Rates: |
|
30, 60, 90, 120, 150, 180, 210, |
|
and 240 BPM |
| | ECG Sensitivity: |
|
50 µV, 100 µV, 200 µV, 0.5 mV, 1 mV, and 2 mV |
|
US-1 tracks primary fetal ECG rates. US-2 tracks secondary fetal activity for either independent “normal” or “twins” simulation, US-2 rate fixed at 140 BPM |
| | Fetal Patterns: |
|
Note: US-1 fetal ECG track these selections. US-2 is in normal pattern, except during TREND #1 selection. |
| | Variability selections (added to fetal ECG) |
|
Absent variability, low variability, mild variability, high variability, severe variability, and long-tern variability |
| | Note: |
|
These patterns repeat and toco channel will perform toco wave selected |
| | Optional mechanical fetal heart |
|
Provides a mechanical interface to the ultrasound transducer; can be connected to either ultrasound channels. This option, due to its power consumption, requires an ac power adapter to be connected. |
|
| Trend #1: Twin Fetal Patterns |
| NORMAL: | | | TACHYCARDIA: | | | BRADYCARDIA: | | | ARRHYTHMIAS: | | | LATE DECELERATION: | | | EARLY DECELERATION: | | | MODERATE DECELER: |
|
Moderate variable deceleration |
| | ACCELERATION #1: | | | ACCELERATION #2: | | | SINUSOIDAL (HIGH): |
|
Sinusoidal pattern, large change |
| | SINUSOIDAL (LOW): |
|
Sinusoidal pattern, small change |
| | SEVERE VAR. DEC. #1: |
|
Severe deceleration wave #2 |
| | SEVERE VAR. DEC. #2: |
|
Severe variable deceleration |
| | PROLONGED DECELERATION: | | | BIPHASIC DECELERATION: | | | EXAGGERATED DECELERATION: | | | NON-UNIFORM DECELERATION: | | | VAR. DECELERATION (U): |
|
Variable deceleration "U" shaped |
| | VAR. DECELERATION TACH: |
|
Variable deceleration with high-rate BPM |
| | VAR. DECELERATION (V): |
|
Variable deceleration "V" shaped |
| | VAR. DECELERATION (POST): |
|
Variable deceleration post exaggerated |
| | VAR. DECELERATION: | | | DECELERATION (POSITION): |
|
Variable deceleration with position changes |
| | LONG DECELERATION: | | | COMPENSATORY ACCELATION: |
|
Compensatory acceleration |
|
| Maternal ECG |
| ECG Static Rates: |
|
60, 80, 100, 120, 140, 160 BPM |
| | ECG Sensitivity: | | | Note: |
|
Pattern selected during Trend #1 selection |
|
| Uterine Activity |
| Note: |
|
Toco waveform selection not available during Trend #1 |
| | EXECUTE WAVEFORM: | | | UTERINE WAVE OFF: | | | ANALOG 0 TO 1 VOLT: |
|
Analog range 0 to 1 V (1 V = 100 mmHg) |
| | ANALOG 0 TO -1 VOLT: |
|
Analog range 0 to -1 V (1 V = 100 mmHg) |
| | UTERINE WAVE 0 TO 25: | | | UTERINE WAVE 0 TO 50: | | | UTERINE WAVE 0 TO 100: | | | SHORT DURATION: |
|
Toco waveform of short duration |
| | NORMAL DURATION: |
|
Normal duration of toco waveform |
| | INCREASED DURATION: |
|
Long duration of toco waveform |
| | UTERINE LEVEL = ZERO: |
|
Zero toco channel (automatic on power up) |
| | UTERINE STATIC + 20: |
|
Increase toco static level by 20 mmHg (0 to 100 mmHg) |
| | INCR. RESTING TONE: | | | COUPLING: | | | TRIPLING: | | | UTERINE PRESSURE SENSITIVITY: | | | Important Notes: |
|
US-1 tracks the fetal ECG rates. |
|
US-2 is the second ultrasound channel with a |
|
normal fetal ECG pattern. |
|
On the fetal and maternal ECG, the fetal ECG |
|
is 1/4 the size of the maternal ECG. |
|
The PS-320 turns on in the following state: |
|
• Fetal ECG static rate @ 150 BPM |
|
• US-1 tracks @ 150 BPM |
|
• US-2 normal pattern |
|
• Pressure sensitivity @ 5 µV/mmHg |
|
• Pressure/Toco set to zero |
|
• Maternal ECG rate @ 80 BPM |
|
• ECG sensitivity @ 1 mV |
|
• Toco wave is normal duration @ 0 to 50 divisions (i.e. 0 to 50 mmHg) |
|
| Controls |
| Display: |
|
2-line x 16-character LCD with keypad |
| | RS232: |
|
Bidirectional interface, 9600 Baud |
|
| General Information |
| Power: |
|
9V battery/battery eliminator; low battery indication set at 6 V |
| | Housing: | | | Dimensions: |
|
6.1 " L x 3.7 " W .x 1.3 " H (15.6 cm L 9.4 cm W x 3.4 cm H) |
| | Weight: | | | Temperature: |
|
59 to 95 ˚F (15 to 35 ˚C), operating; 32 to 122 ˚F (0 to 50 ˚C), storage |
|
|
| Model Name | Description | Price List | | PS320 | Fetal Simulatorspan
Includes:
- Electronic-version users manual (CD-ROM)
- Printed-version Users Manual (MANUAL)
- Battery eliminator universal 90 to 240 V (BE-UNVSL-IEC320C14)
- 9 V battery
|
| | Kit #1: GE Corometrics | PS320 Fetal Monitoring Kit, GE Corometrics
Includes:
- PS320 Fetal Simulator (PS320, includes Operator Manual MANUAL, Battery Eliminator BE-UNVSL-IEC320C14, and 9 V Battery)
- MFH-1 Mechanical Fetal Heart Probe –includes Fetal Heart Cable (17156)
- Soft-Sided Carrying Case for kit (ACCU LUNG CASE)
- RS-232 Cable, PS320/420 (17291)
- Corometrics Ultrasound Simulation Cable (17483)
- Corometrics IUP Simulation Cable (17481)
- Corometrics TOCO Simulation Cable (17486)
- Ultrasound Simulation Cable Series 120 (PN 4021537)
|
PS320 Fetal Simulation cable for GE Corometrics Fetal monitors
|
|
Corometrics Series
|
Ultrasound PN 4021537
|
Ultrasound PN 2462484 (#17483)
|
IUP PN 2462469 (#17481)
|
TOCO PN 2462519(#17486)
|
|
116
|
No
|
Yes
|
Yes
|
Yes
|
|
118
|
No
|
Yes
|
Yes
|
Yes
|
|
120
|
Yes
|
No
|
Yes
|
No
|
|
170
|
No
|
Yes
|
No
|
Yes
|
|
250
|
No
|
Yes
|
Yes
|
Yes
|
| |
| | Kit #2: Hewlett-Packard, Agilent, Philips Medical | PS320 Fetal Monitoring Kit, Series 50/8040 Philips Medical, Hewlett-Packard, Agilent
Includes:
- PS320 Fetal Simulator (PS320, includes Operator Manual MANUAL, Battery Eliminator BE-UNVSL-IEC320C14, and 9 V Battery)
- MFH-1 Mechanical Fetal Heart Probe –includes Fetal Heart Cable (17156)
- Soft-Sided Carrying Case for kit (ACCU LUNG CASE)
- RS-232 Cable, PS320/420 (17291)
- HP/AG/PHILIPS (50 SERIES) Ultrasound Simulation Cable PS320 (17482)
- HP/AG/PHILIPS (50 SERIES) EXT TOCO Simulation Cable PS320 (17484)
- HP/AG/PHILIPS (50 &8040 SERIES) IUP Simulation Cable PS320 (17487)
- HP (8040 SERIES) Ultrasound Simulation Cable PS320 (17488)
- HP (8040 SERIES) EXT TOCO Simulation Cable PS320 (17489)
|
| | Kit #3: Oxford Medical | PS320 Fetal Monitoring Kit, Oxford Medical
Includes:
- PS320 Fetal Simulator (PS320, includes Operator Manual MANUAL, Battery Eliminator BE-UNVSL-IEC320C14, and 9 V Battery)
- MFH-1 Mechanical Fetal Heart Probe –includes Fetal Heart Cable (17156)
- Soft-Sided Carrying Case for kit (ACCU LUNG CASE)
- RS-232 Cable, PS320/420 (17291)
- OXFORD IUP Simulation Cable (17492)
- OXFORD Ultrasound Simulation Cable 2 MHz (blue) PS320 (17491)
- OXFORD Ultrasound Simulation Cable 1.5 MHz (yellow) PS320 (17490)
|
|
| Batteries, Chargers, Power Supply | | BE-UNVSL-IEC320C14 | BE-UNVSL-IEC320C14
, Battery eliminator 100 to 240 VAC(part#2647372) |
| Cables | | 17484 | 17484,Agilent 50 Series TOCO - External cable | | 17487 | 17487,Agilent 50/8040 series IUP cable
| | 17483 | 17483,Corometrics - Ultrasound cable | | 17481 | 17481,Corometrics TOCO - External cable | | 17486 | 17486,Corometrics TOCO - IUP cable | | 17489 | 17489,HP (8040 Series) ext TOCO simulation cable | | 17488 | 17488,HP (8040 Series) ultrasound simulation cable | | 17156 | 17156,Mechanical fetal heart cable
| | 17492 | 17492,Oxford IUP cable | | 17490 | 17490,Oxford ultrasound simulation cable 1.5 MHz (yellow)(Part#2462555) | | 17491 | 17491,Oxford ultrasound simulation cable 2.0 Mhz (blue | | 17482 | 17482,Philips 50 Series - Ultrasound cable | | 17291 | 17291,RS232-cable |
| Carrying Cases | | 17192 | 17192, Carrying Case, double pocket, soft |
| Miscellaneous Accessories | | MFH-1 | MFH-1, Mechanical fetal heart probe |
|
|
|
|
|
|
|
|
|
 | | Search Fluke Biomedical | | | | | | | | | | | | | | | | | | | | | | | | Find Resources | | | | | | | | |
|
|
|
| |  | |
|