FlukeBioMed
Home  |  Products  |  Patient Simulators  |  
Contact us

PS320 Fetal Simulator

 PS320

PS320 Fetal Simulator

The PS320 Fetal Simulator simulates fetal and maternal ECG as well as uterine activity to test and troubleshoot fetal electronic monitors and to train clinical staff.

The unit is battery operated and small enough to fit in a pocket so mobile technicians and clinical instructors can take it anywhere.

The PS320 simulates several fetal parameters including twins, and it simulates a wide range of clinical scenarios for training labor-and-delivery staff in how to recognize normal and abnormal responses. An optional mechanical heart creates fetal heart sounds for testing fetal monitors' ultrasound cables and transducers.

PS320 offers an easy user interface providing a 2 x 16-character LCD display with adjustable contrast. The unit operates on a 9-volt battery with low-battery monitoring or functions with the supplied battery eliminator.

*The PS320 is not compatible with the Phillips Avalon fetal monitor.

Key Features

  • Mechanical heart for ultrasound simulation
  • TOCO simulation (External or IUP)
  • Ultrasound simulation (including twins)
  • Maternal ECG simulation
  • Fetal ECG (tracks ultrasound #1)
  • Internal (DECG) and external fetal ECG
  • Uterine-activity selections
  • Fetal beat-to-beat variability
  • Periodic and non-periodic fetal ECG changes
  • Arrhythmia selections
  • Compact, lightweight, pocket-size plastic housing
  • Battery operated with status indications
  • Special kits available with all required accessories and cables to test fetal monitors for specified manufacturers

Specifications

Fetal ECG
Static Rates:
 30, 60, 90, 120, 150, 180, 210,
 and 240 BPM
ECG Sensitivity:
 50 µV, 100 µV, 200 µV, 0.5 mV, 1 mV, and 2 mV
 US-1 tracks primary fetal ECG rates. US-2 tracks secondary fetal activity for either independent “normal or “twins simulation, US-2 rate fixed at 140 BPM
Fetal Patterns:
 Note: US-1 fetal ECG track these selections. US-2 is in normal pattern, except during TREND #1 selection.
Variability selections (added to fetal ECG)
 Absent variability, low variability, mild variability, high variability, severe variability, and long-tern variability
Note:
 These patterns repeat and toco channel will perform toco wave selected
Optional mechanical fetal heart
 Provides a mechanical interface to the ultrasound transducer; can be connected to either ultrasound channels. This option, due to its power consumption, requires an ac power adapter to be connected.

 

Trend #1: Twin Fetal Patterns
NORMAL:
 Normal pattern
TACHYCARDIA:
 Tachycardia pattern
BRADYCARDIA:
 Bradycardia pattern
ARRHYTHMIAS:
 Arrhythmia pattern
LATE DECELERATION:
 Late deceleration
EARLY DECELERATION:
 Early deceleration
MODERATE DECELER:
 Moderate variable deceleration
ACCELERATION #1:
 Acceleration wave #1
ACCELERATION #2:
 Acceleration wave #2
SINUSOIDAL (HIGH):
 Sinusoidal pattern, large change
SINUSOIDAL (LOW):
 Sinusoidal pattern, small change
SEVERE VAR. DEC. #1:
 Severe deceleration wave #2
SEVERE VAR. DEC. #2:
 Severe variable deceleration
PROLONGED DECELERATION:
 Prolonged deceleration
BIPHASIC DECELERATION:
 Biphasic deceleration
EXAGGERATED DECELERATION:
 Exaggerated deceleration
NON-UNIFORM DECELERATION:
 Non-uniform deceleration
VAR. DECELERATION (U):
 Variable deceleration "U" shaped
VAR. DECELERATION TACH:
 Variable deceleration with high-rate BPM
VAR. DECELERATION (V):
 Variable deceleration "V" shaped
VAR. DECELERATION (POST):
 Variable deceleration post exaggerated
VAR. DECELERATION:
 Variable deceleration
DECELERATION (POSITION):
 Variable deceleration with position changes
LONG DECELERATION:
 Long deceleration
COMPENSATORY ACCELATION:
 Compensatory acceleration

 

Maternal ECG
ECG Static Rates:
 60, 80, 100, 120, 140, 160 BPM
ECG Sensitivity:
 0.5 mV, 1 mV, and 2 mV
Note:
 Pattern selected during Trend #1 selection

 

Uterine Activity
Note:
 Toco waveform selection not available during Trend #1
EXECUTE WAVEFORM:
 Start toco waveform
UTERINE WAVE OFF:
 Stop toco waveform
ANALOG 0 TO 1 VOLT:
 Analog range 0 to 1 V (1 V = 100 mmHg)
ANALOG 0 TO -1 VOLT:
 Analog range 0 to -1 V (1 V = 100 mmHg)
UTERINE WAVE 0 TO 25:
 Range of toco waveform
UTERINE WAVE 0 TO 50:
 Range of toco waveform
UTERINE WAVE 0 TO 100:
 Range of toco waveform
SHORT DURATION:
 Toco waveform of short duration
NORMAL DURATION:
 Normal duration of toco waveform
INCREASED DURATION:
 Long duration of toco waveform
UTERINE LEVEL = ZERO:
 Zero toco channel (automatic on power up)
UTERINE STATIC + 20:
 Increase toco static level by 20 mmHg (0 to 100 mmHg)
INCR. RESTING TONE:
 Resting tone increases
COUPLING:
 2 close toco waves
TRIPLING:
 3 close toco waves
UTERINE PRESSURE SENSITIVITY:
 5 or 40 µV on power up
Important Notes:
 US-1 tracks the fetal ECG rates.
 US-2 is the second ultrasound channel with a
 normal fetal ECG pattern.
 On the fetal and maternal ECG, the fetal ECG
 is 1/4 the size of the maternal ECG.
 The PS-320 turns on in the following state:
 • Fetal ECG static rate @ 150 BPM
 • US-1 tracks @ 150 BPM
 • US-2 normal pattern
 • Pressure sensitivity @ 5 µV/mmHg
 • Pressure/Toco set to zero
 • Maternal ECG rate @ 80 BPM
 • ECG sensitivity @ 1 mV
 • Toco wave is normal duration @ 0 to 50 divisions (i.e. 0 to 50 mmHg)

 

Controls
Display:
 2-line x 16-character LCD with keypad
RS232:
 Bidirectional interface, 9600 Baud

 

General Information
Power:
 9V battery/battery eliminator; low battery indication set at 6 V
Housing:
 Plastic case
Dimensions:
 6.1 " L x 3.7 " W .x 1.3 " H (15.6 cm L 9.4 cm W x 3.4 cm H)
Weight:
 0.9 lb (0.4 kg)
Temperature:
 59 to 95 ˚F (15 to 35 ˚C), operating; 32 to 122 ˚F (0 to 50 ˚C), storage

*The PS320 is not compatible with the Phillips Avalon fetal monitor.

Model NameDescription
PS320Fetal Simulator

Includes:

  • Electronic-version users manual (CD-ROM)
  • Printed-version Users Manual (MANUAL)
  • Battery eliminator universal 90 to 240 V (BE-UNVSL-IEC320C14)
  • 9 V battery

Kit #1: GE CorometricsPS320 Fetal Monitoring Kit, GE Corometrics

Includes:

  • PS320 Fetal Simulator (PS320, includes Operator Manual MANUAL, Battery Eliminator BE-UNVSL-IEC320C14, and 9 V Battery)
  • MFH-1 Mechanical Fetal Heart Probe –includes Fetal Heart Cable (17156)
  • Soft-Sided Carrying Case for kit (ACCU LUNG CASE)
  • RS-232 Cable, PS320/420 (17291)
  • Corometrics Ultrasound Simulation Cable (17483)
  • Corometrics IUP Simulation Cable (17481)
  • Corometrics TOCO Simulation Cable (17486)
  • Ultrasound Simulation Cable Series 120 (PN 4021537)

PS320 Fetal Simulation cable for GE Corometrics Fetal monitors

Corometrics Series

Ultrasound PN 4021537

Ultrasound PN 2462484 (#17483)

IUP PN 2462469 (#17481)

TOCO PN 2462519(#17486)

116

No

Yes

Yes

Yes

118

No

Yes

Yes

Yes

120

Yes

No

Yes

No

170

No

Yes

No

Yes

250

No

Yes

Yes

Yes

 
Kit #2: Hewlett-Packard, Agilent, Philips MedicalPS320 Fetal Monitoring Kit, Series 50/8040 Philips Medical, Hewlett-Packard, Agilent

Includes:

  • PS320 Fetal Simulator (PS320, includes Operator Manual MANUAL, Battery Eliminator BE-UNVSL-IEC320C14, and 9 V Battery)
  • MFH-1 Mechanical Fetal Heart Probe –includes Fetal Heart Cable (17156)
  • Soft-Sided Carrying Case for kit (ACCU LUNG CASE)
  • RS-232 Cable, PS320/420 (17291)
  • HP/AG/PHILIPS (50 SERIES) Ultrasound Simulation Cable PS320 (17482)
  • HP/AG/PHILIPS (50 SERIES) EXT TOCO Simulation Cable PS320 (17484)
  • HP/AG/PHILIPS (50 &8040 SERIES) IUP Simulation Cable PS320 (17487)
  • HP (8040 SERIES) Ultrasound Simulation Cable PS320 (17488)
  • HP (8040 SERIES) EXT TOCO Simulation Cable PS320 (17489)

Kit #3: Oxford MedicalPS320 Fetal Monitoring Kit, Oxford Medical

Includes:

  • PS320 Fetal Simulator (PS320, includes Operator Manual MANUAL, Battery Eliminator BE-UNVSL-IEC320C14, and 9 V Battery)
  • MFH-1 Mechanical Fetal Heart Probe –includes Fetal Heart Cable (17156)
  • Soft-Sided Carrying Case for kit (ACCU LUNG CASE)
  • RS-232 Cable, PS320/420 (17291)
  • OXFORD IUP Simulation Cable (17492)
  • OXFORD Ultrasound Simulation Cable 2 MHz (blue) PS320 (17491)
  • OXFORD Ultrasound Simulation Cable 1.5 MHz (yellow) PS320 (17490)


Test Leads, Probes and Clips
MFH-1MFH-1 Mechanical fetal heart probe (part# 2651757)
Includes fetal heart cable

Batteries, Chargers, and Power Supply
BE-UNVSL-IEC320C14BE-UNVSL-IEC320C14 Battery Eliminator (part# 2647372)
Regulated, 115 VAC/9 VDC
for PS320, PS410, PS420, DPM4

Cables
1748917489 HP/Ag/Philips 8040 Series TOCO simulation cable (12-pin) (part# 2462543) 1.75m
1748617486, Corometrics TOCO - IUP simulation cable for MPS450 PS420, PS320 (external) (12-pin) (part# 2462519)
Mini-DIN Male to Corometrics, 1.75m
1748317483 Corometrics 118 ultrasound simulation cable for PS320 (12-pin) (part# 2462484)
Mini-DIN Male to Corometrics, 1.75m
1748117481 Corometrics TOCO IUP simulation cable- External (part# 2462469)
Mini-DIN Male to Corometrics, 1.75m
1748717487 TOCO IUP simulation cable HP/Ag/Philips (50 & 8040 Series) for PS320/PS420/MPS450 (12-pin) (part# 2462528)
1748417484 HP, Philips, Agilent 50 Series TOCO - External cable for PS320 (part# 2462491)
Mini-DIN Male to Agilent style connector, 1.75m
1748217482 HP/Ag/Philips 50 Series - Ultrasound simulation cable for PS320 (12-pin) (part# 2462478)
Mini-DIN Male to HP style, 5 ft
1748817488 HP/Ag/Philips 8040 Ultrasound simulation cable for PS320 (12-pin) (part# 2462537)
Mini-DIN male to Agilent style connector, 1.75m
1715617156 MFH-1 Mechanical fetal heart cable (part# 2462123)
1749017490, Oxford 1.5 MHz ultrasound simulation cable (yellow)(12-pin) (part# 2462555)
for PS320 Fetal Simulator, 1.75m
1749117491 Oxford 2.0MHz ultrasound simulation cable (blue) (12-pin) (part# 2462562)
1749217492 Oxford IUP simulation cable for PS320 (12-pin) (red) (part# 2462570)
1729117291 RS232 cable (part# 2462217) for PS320/PS420
1.5m, 6 pin Mini Din Male to 9 pin D-Sub Female

Carrying Cases
1719217192 Carrying case (Soft) Double Pocket (part# 2462177)

Miscellaneous Accessories
26400142640014 PS320 case with ECG posts (part# 3539794)
PS320 ECG Posts are only sold at this assembly level.
3113821Battery cover (part# 3113821)
for PS420, PS410, PS320, DPM4

Service Packages
Care PlansRequest a quote
Biomedical Equipment  |  Patient Simulator  |  Safety Analyzer  |  Dosimeter  |  Radiation Detector
Search Fluke Biomedical
ResourcesProducts
Find Resources
INCU incubator analyzers by Fluke Biomedical

PS320 Fetal Simulator Data Sheet


Data Sheet
Fiche Technique
Technische Daten
Datos Técnicos
Caratteristiche Tecniche
Teknik Bilgiler
Dados Técnicos
データシート
数据表
Fluke Biomedical PS320 Cable Selection Guide

Follow us Fluke Biomedical on Facebook Fluke Biomedical on Twitter Fluke Biomedical on YouTube Fluke Biomedical on LinkedIn Home | Site Map | Fluke Corporation | Terms and Conditions | Privacy Statement | Disclaimer © 1995 - 2014 Fluke Corporation