IDA-1S Infusion Device Analyzer

Test a wide variety of infusion pumps quickly and accurately.

Get pricing and/or product demo

Watch video now »
IDA-1S infusion device analyzer
IDA-1S infusion device analyzer
IDA-1S infusion pump analyzer
IDA-1S infusion device tester
IDA-1S infusion pump tester

IDA-1S Infusion Device Analyzer

Ensure infusion pumps are tested accurately and quickly with real time results using the IDA-1S One-Channel Infusion Device Analyzer, the newest member of the Fluke Biomedical IDA family. The IDA-1S infusion device analyzer is a portable, battery-operated instrument that allows for speedy verification when testing infusion device performance.

The IDA-1S infusion pump tester measures the flow rate and volume delivered, and the pressure generated in occlusions or blockages of the fluid line. The IDA-1S infusion device tester is based on sophisticated measurement technology trusted by biomedical professionals around the world.

It is easy to set up and requires little or no training to use. The IDA-1S infusion device analyzer can be used to test a wide variety of infusion pumps and an auto-start feature simplifies syringe pump testing and other tests that have long startup times.

  • Integrated carrying handle and lightweight (2.7 lb) for easy portability
  • Battery powered with up to 10 hours of continuous operation for on-the-go operation
  • LCD touch screen for ease of use
  • Average and instantaneous flow measurement
  • Occlusion pressure measurements to 45 psi
  • Maximize accuracy with Autostart mode enabling unit to begin testing only when fluid is detected
  • Compatible with a wide variety of infusion pumps
  • Based on technology that is proven and trusted worldwide
  • On-board memory allows test results storage instantly
  • Hydrograph graphical software to control unit,display results and print results via PC
  • Global sales, service and support

Specifications

 

Flow rate measurement
Technique
 Flow is calculated by measuring volume over time
Range
 0.5 ml/h to 1000 ml/h
Accuracy
 1 % of reading ±1 LSD for flows of 16 ml/h to 200 ml/h for volumes over 20 ml, otherwise 2 % of reading ±1 LSD for volumes over 10 ml under laboratory conditions.
Max test duration
 10 hours on battery
Volume measurement
Technique
 Volume measured directly by the measuring module in minimum sample sizes of 60 microliters
Range
 0.06 ml to 999 ml
Accuracy
 1 % of reading ±1 LSD for flow rates of 16 ml/h to 200 ml/h for volumes over 20 ml. Otherwise 2 % of reading ±1 LSD for volumes over 10 ml under laboratory conditions.
Max test duration
 10 hours on battery
Pressure measurement
Technique (Occlusion test)
 Direct measurement of pressure at the inlet port
Range
 0 psi to 45 psi or equivalent in mmHg and kPa
Accuracy
 1 % of full scale ±1 LSD under laboratory conditions
Max test duration
 30 minutes
General specifications
Storage of results
 Test results stored for later viewing, printing or transfer to PC. Typical practical capacity: 100 tests
Power Down
 The results of tests in progress will be saved in the case of accidental power down
Computer control
 The Product can be fully controlled from a PC using HydroGraph V3 software for IDA-1S
Battery power
 4 x Panasonic HHR210AB NiMh 2000 mAh batteries
Charger
 •Operating voltage range: 100V ac to 240V ac
 •Supply frequency: 50Hz / 60 Hz
 •Supply power: <20 VA
Size (HxWxD)
 30 cm x 17 cm x 10 cm
 (12 in x 8 in x 4 in)
Weight
 ~1.2 kg (2.7 lbs)
Safety
 IEC 61010-1: Overvoltage category II, Pollution Degree 2
Electromagnetic environment
 IEC 61326-1: Basic
Emissions classification
 IEC CISPR 11: Group 1, Class A. Group 1 have intentionally generated and/or use conductively coupled radio-frequency energy which is necessary fpr the internal functioning of the equipment itself. Class A equipment is suitable for use in nondomestic locations and/or directly connected to a low-voltage power supply network
FCC
 CFR47: Class A Part 15 subpart B
Electromagnetic compatibility
 Applies to use in Korea only. Class A: Equipment (Industrial Broadcasting and Communication Equipment)¹
Altitude
 0 m to 2000 m (0 ft to 6500 ft)
Temperature
Operating
 15 °C to 30 °C (59 °F to 86 °F)
Storage
 -20 °C to +40 °C (-4 °F to +104 °F) when drained of all liquid
Humidity
 10 % to 90 % non-condensing

¹ This product meets requirements for industrial (Class A) electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and is not to be used in homes.

Model NameDescription
IDA-1SIDA-1S One-channel Infusion Device Analyzer
Included accessories:
  • Hydrograph Software and Users Manual
  • 20 ml syringe
  • 3-way plastic Luerlock
  • Drain tube (1 ml)
  • Micro-90 bottle (225 ml)
  • USB data transfer cable
  • Power supply accessory (country-specific)
  • Power cord
  • NiMH replacement battery

Cables
USB CableUSB Cable
• for ProSim 2 & 3 (part# 1671807)
• for Impulse 6000D/7000DP (part# 1626219)
• for IDA-1S (part# 1626219)
• for IDA-5 (part# 4354452)

Miscellaneous Accessories
20ML SyringeSyringe, 20ML Syringe (Part #435014)
48457FG48457FG Luerlock 3-way t-valve
(part# 2391750) for IDA-5, IDA-1S, and IDA4 Plus
Drain tubeIDA-5 DRAIN TUBE,5-FT PLASTIC DRAIN LINE IDA-5
(Part #4354429)
MICRO-90MICRO-90 bottle 225ml (part# 4541948)