Fluke Biomedical - Trusted for the measurements that matter to medical industry professionals.
Home  |  Products  |  Gas Flow Analyzers  |  

VT900 Gas Flow Analyzer

The VT900 provides the highest accuracy on the market for testing gas flow and respiratory medical equipment.

VT900 Gas Flow Analyzer
VT900 Gas Flow Analyzer
VT900 Gas Flow Tester
VT900 Gas Flow Tester

VT900 Gas Flow Analyzer

Accurate

The VT900 is Fluke Biomedical’s high-accuracy premium gas flow analyzer. The single, full range ±300 lpm air flow channel offers built-in oxygen, temperature and humidity measurements to help streamline testing and automatically compensate for environmental conditions. The VT900 features an external trigger input and special ultra-low flow and ultra-low pressure ports. These ultra-low flow and ultra-low pressure ports allow the highest accuracy for devices requiring crucial low volume and pressure testing such as anesthesia machines and flow meters. Designed and tested to world renown Molbloc-L calibration specifications ensures traceability to global regulatory standards with reliable measurements you can count on.

Easy-to-use

The VT900 offers a large 7 (17.8 cm) touch screen display, allowing you to view multiple measurements at once, and quickly access menu options. Review results in real-time with either color graphs or numerical data. The global user interface makes operating this device straightforward and uncomplicated.

Traceable

The large on-board memory of the VT900 allows both short and long term recording and storing of test data. Transfer data via USB to a PC and upload the generated test file to your CMMS system for simple reporting. This device can be easily adapted to specific testing needs. With the ability to create custom profiles and the capacity to take remote commands for automated testing, the VT900 helps to decrease risk and increase efficiency.

Portable

Weighing only 3.6 lb (1.6 kg), this compact, all-in-one device is highly portable. The snap-in carrying handle/shoulder strap and rugged design allows you to easily test on-the-go, while its small unit size and bale (kick stand) allows comfortable viewing for benchtop testing. With AC/DC power options and an 8-hour battery life, this tester is perfect for laboratory, clinical or field environments where AC power may not be available.

VT900 Gas Flow Analyzer

Key benefits and features

  • Streamline your testing procedure, reduce errors and quicken your test time with the ability to create customized test profiles
  • Avoid confusion and ensure accuracy with one-channel, full range air flow functionality
  • Reduce testing time with built-in line sensors which automatically test humidity, temperature and oxygen while compensating for atmospheric pressure and environmental conditions
  • Ensure patient safety with ultra-low flow and ultra-low pressure anesthesia and flow meter testing
  • Have confidence that your measurements comply to global regulatory standards and adhere to SI units of measurement with the Molbloc-L calibration system.
  • Easily transport and store the lightweight (3.6lb/1.6kg), all-in-one device – no extra modules for different tests
  • Have more control over your testing by selecting your own trigger point with the external trigger input
  • Operate on-the-go, all day with 8 hours of battery life

VT900 Gas Flow Analyzer

Features

Battery life hours

8 hrs

Charge time in hours

5 hrs, typical

Memory

internal memory

Connection type

USB, Micro-B device port

Weight

3.6 lb (1.6 kg)

Display

7 in (17.8 cm)

Single full-range channel

Ultra-low flow ports

±750 ml/min

Ultra-low pressure port

0 to 10 mbar

Flow

Full range flow channel (includes both low and high flow)

Range

±300 slpm

Accuracy (air)

1.7 % or 0.04 slpm

Ultra-low flow channel

Range

±750 ml/min

Accuracy (air)

±1.7 % or 0.01 slpm

Volume

Range

±100 l

Accuracy

±1.75 % or 0.02 l

Pressure

High pressure

Range

-0.8 to 10 bar

Accuracy

±1 % or ±0.007 bar

Differential low pressure

Range

±160 mbar

Accuracy

±0.5 % or ±0.1 mbar

Ultra-low pressure

Range

0 to 10 mbar

Accuracy

±1 % or ±0.01 mbar

Airway pressure

Range

±160 mbar

Accuracy

±0.5 % or ±0.1 mbar

Barometric pressure

Range

550 to 1240 mbar

Accuracy

±1 % or ±5 mbar

Other

Temperature

Range

0 to 50 °C

Accuracy

±0.5 °C

Resolution

0.1 °C

Humidity

Range

0 to 100 % RH

Accuracy

±3 % RH (20 to 80 % RH)
±5 % RH (20< or >80 % RH)

Oxygen

Range

0 to 100 %

Accuracy

±1 %

Breath parameters

Inspiratory tidal volume range

0 to 60 l

Inspiratory tidal volume accuracy

±1.75 % or 0.02 l

Expiratory tidal volume range

0 to 60 l

Expiratory tidal volume accuracy

±1.75 % or 0.02 l

Minute volume range

0 to 100 l

Minute volume accuracy

±1.75 % or 0.02 l

Breath rate range

1 to 1500 bpm

Breath rate accuracy

±1 %

Inspiratory to expiratory time ratio (I:E) range

1:300 to 300:1

Inspiratory to expiratory time ratio (I:E) accuracy

±2 % or 0.1

Peak inspiratory pressure (PIP) range

±160 mbar

Peak inspiratory pressure (PIP) accuracy

±0.75 % or 0.1 mbar

Inspiratory pause pressure range

±160 mbar

Inspiratory pause pressure

±0.75 % or 0.1 mbar

Mean airway pressure range

±160 mbar

Mean airway pressure accuracy

±0.75 % or 0.1 mbar

Positive end expiratory pressure (PEEP) range

±160 mbar

Positive end expiratory pressure (PEEP) accuracy

±0.75 % or 0.1 mbar

Lung compliance range

0 to 1000 ml/mbar

Lung compliance accuracy

±3 % or 0.1 ml/mbar

Inspiratory time range

0 to 60 s

Inspiratory time accuracy

0.02 s

Inspiratory hold time range

0 to 60 s

Inspiratory hold time accuracy

1 % or 0.1 s

Expiratory time range

0 to 90 s

Expiratory time accuracy

0.5 % or 0.01 s

Expiratory hold time range

0 to 90 s

Expiratory hold time accuracy

0.02 s

Peak expiratory flow range

±300 lpm

Peak expiratory flow accuracy

±1.7 % or 0.04 lpm

Peak inspiratory flow range

±300 lpm

Peak inspiratory flow accuracy

±1.7 % or 0.04 lpm

Environmental

Operating temp

10 °C to 40 °C

Storage temp

-20 °C to 60 °C

Operating humidity

10 to 90 % non-condensing

Storage humidity

5 to 95 % non-condensing

Gas corrections

Gas types

ATP (ambient temp/pressure, actual humidity)

Air

ATPD (ambient temp/pressure, dry)

Nitrogen (N2)

ATPS (ambient temp/pressure, saturated)

Nitrous Oxide (N2O)

STP20 (20 °C temp/pressure 760 mmHg, actual humidity)

Carbon Dioxide (CO2)

STP21 (21 °C temp/pressure 760 mmHg, actual humidity)

Oxygen (O2)

STPD0 (0 °C temp/pressure 760 mmHg, dry)

Argon

STPD20 (20 °C temp/pressure 760 mmHg, dry)

Heliox (21 % O2, 79% He)

STP or STPD21 (21 °C temp/pressure 760 mmHg, dry)

Oxygen/Nitrogen

BTPS (body temp 37 °C/ambient pressure 760 mmHg, saturated)

Oxygen/Nitrous Oxide

BTPD (body temp 37 °C/ambient pressure 760 mmHg, dry)

Oxygen/Helium

Model NameDescription
VT900, GAS FLOW ANALYZER

VT900,GAS FLOW ANALYZER (Part #4920333)
Includes the following:

  • Certificate of Calibration with test data
  • Carrying case
  • Carrying handle and shoulder strap
  • USB serial cable
  • Tilt stand
  • AC power adapter
  • Bacterial filter
  • 1.2 m (4 ft) silicon tubing (2)
  • 22 mm ID x 22 mm ID tubing adapters (2)
  • 22 mm OD x 22 mm OD tubing adapters (2)
  • Tapered 15 mm OD x 33 mm OD tubing adapters (2)
  • Flexible 15 mm ID x 22 mm ID tubing adapters (2)
  • DISS hand tight nut/nipple to 6.4 mm (1/4 in ) ID hose barb adapter (1)

Gas Flow Analyzers
ACCU LUNG Test LungACCU LUNG Portable Precision Test Lung
ACCULUNG IIAccuLung II Adjustable Portable Precision Test Lung
(part# 4281291)

Service Packages
Care PlansRequest a quote
Related Products
View related products
 
 VT PLUS HF Gas Flow Analyzer »
 
 VT305 Gas Flow Analyzer »
 
 VT650 Gas Flow Analyzer »
Follow us Fluke Biomedical on Facebook Fluke Biomedical on Twitter Fluke Biomedical on YouTube Fluke Biomedical on LinkedIn Home | Site Map | Terms of Use | Terms and Conditions | Privacy Policy | Disclaimer © 1995 - 2018 Fluke Corporation