The TNT 12000 X-Ray Test Tools system is the newest and most comprehensive family of instruments available for assuring quality and safety of diagnostic x-ray imaging systems. With selection of all-in-one-exposure solid-state detector, dosimeter, ion chambers, optional mA/mAs invasive shunt or non-invasive clamp device, and choice of handheld display or laptop interface (both completely wireless), the TNT 12000 X-Ray Test Tools provide state-of-the-art solutions for any x-ray safety and analysis test protocol. A long battery life ensures uninterrupted operation all day.
The 12000WD solid-state detector sets up in seconds and measures kVp, radiation dose, dose rate, time, and half value layer (HVL) in a single exposure. The companion TNT 12000 DoseMate dosimeter and ion chambers provide precision dose and dose-rate measurement of radiographic, dental, fluoroscopic, and CT imaging systems. The wireless ZigBee® interface allows for quick testing and reporting, and at less than 1 mW power is preferred over Bluetooth® in medical settings where interference with delicate patient monitoring and treatment equipment may be a concern.
TNT 12000WD Wireless Detector
Featuring all-in-one-exposure measurements and ZigBee wireless communication combined with the rugged, reliable, and accurate design that is a Fluke trademark, the TNT 12000WD is a new breed of non-invasive x-ray test tool. Its compact design enhances portability and wireless operation ensures setup in seconds.
Because the TNT 12000WD measures all parameters with every exposure, there is no need for complicated menu selection, further enhancing user productivity. TNT 12000WD always defaults to the last use when powered on, so when used often for repetitive procedures it is truly a one-button (power on) solution. TNT 12000WD has the expanded functionality needed for modern applications and can be managed with minimum keystrokes. Users can identify and select custom measurement protocols and save them for future use. Full test automation and documentation software is available for TNT 12000WD, creating the advantage of accurate, repeatable testing processes.
TNT 12000 DoseMate dosimeter
The TNT 12000 DoseMate with ion-chambers provides the precision dose measurements needed for absolute dose measurement integrity. Offering customizable measurement protocols, the DoseMate dosimeter provides comprehensive, repeatable testing for radiographic, dental, fluoroscopic, and CT imaging systems in just a few keystrokes. The TNT 12000 DoseMate is compatible with existing Triad and NERO external ion chambers, offering the newest dosimeter technology with minimum investment.
mA/mAs measurement
Integrated mA/mAs measurement capability is available with both the TNT 12000WD and the TNT 12000 DoseMate. A mA/mAs-standard invasive shunt or optional non-invasive clamp device are available to meet the challenges of mAs measurement with all types of x-ray systems. ZigBee wireless interface delivers measured results to the user’s choice of handheld display or laptop, integrating all QA measurements on a single readout.
TNT 12000D Wireless Display
The TNT 12000D Wireless Display has been specifically designed with user productivity in mind and displays all measured results, as well as identification information, battery level, calibration date and more. ZigBee wireless interface from the test device offers targeted communication to the mating display for speedy results reading up to 100 feet from the test device. Wireless communication and intelligent device interface allows the user to perform multiple exposures without leaving the radiation-shielded area, thus allowing complete test protocols to be performed with minimum travel between operator and test device. Low power requirements for ZigBee interface allow for extensive battery life as compared to Bluetooth test devices, which require frequent recharging from inefficient power drain.
Ansur Test Automation Software
Test automation software allows preventative maintenance routines to be tightly controlled and consistently reported while directing users through prescribed test protocols with simple and efficient instruction. Test standardization minimizes human error and alleviates training requirements with step-by-step instruction. Photos, drawings and instructional visual aids included in test protocols negate the need to carry bulky service manuals for complicated PM procedures. Professional electronic reporting and database/CMMS interface allows for total PM and service data management.
|
- Choice of all-in-one detector, dosimeter, integral ma/mAs and handheld display or user’s own laptop interface.
- 100 % ZigBee wireless operation
- Compact handheld design
- Sets up in seconds
- Displays all values in one shot (all-in-one detector)
- Simple user interface
- Unbeatable ruggedness for long-term reliability
- 40 kHz kV sampling rate to ensure accuracy with the most difficult applications (TNT 12000WD)
- Global support network delivering prompt service and peace of mind to Fluke Biomedical customers worldwide
|
|
|
TNT 12000 X-Ray Test Tools
| Physical specifications |
| Display | | | Size (WxDxH) |
| TNT 12000D Display |
6.1 in x 4.5 I x 1.6 in |
| TNT 12000WD Wireless Detector |
6.7 in x 4.5 in x 1.75 in |
| TNT 12000 DoseMate Dosimeter |
6.7 in x 4.5 in x 1.75 in |
| | Weight |
| TNT 12000D Display |
0.422 kg (0.93 lb) |
| TNT 12000WD Wireless Detector |
1.5 lb |
| TNT 12000 DoseMate Dosimeter |
1.1 lb |
|
| Electrical specifications |
| Battery |
| Type |
Lithium-Ion, 3.7 V, 4000 mAh |
| Charge time |
Approximately 5 hr |
| Operating duration |
Approximately 8 hr minimum |
| | AC adapter |
| Input voltage |
100 V ac to 240 V ac |
| Input frequency |
50/60 Hz |
| Input current |
0.5 A (rms) |
| Output voltage |
6 V dc |
|
| Environmental specifications |
| Operating temperature |
|
0 ºC to 35 ºC (32 ºF to 122 ºF) |
| | Storage temperature |
|
-35 ºC to 50 ºC (-31 ºF to 122 ºF) |
| | Operating humidity |
|
20 % to 80 % RH (non-condensing) |
|
| Measurements |
| kVp |
| Units |
kVp Average (average of peaks during a specified interval) |
|
kVp Max (highest peak during a specified interval) |
|
PPV (peak practical voltage) |
| Ranges |
Radio/Fluoro modes: 40 kV to 150 kV |
|
Mammo modes:
Mo/Mo: 22 kV to 35 kV (standard calibration)
Rh/Rh: 25 kV to 49 kV (optional calibration)
Mo/Rh: 22 kV to 40 kV (optional calibration)
Mo/Al: 22 kV to 49 kV (optional calibration)
Rh/Al: 22 kV to 49 kV (optional calibration)
W/Rh: (optional calibration)
W/Ag: (optional calibration) |
| Resolution |
0.1 kV |
| Accuracy |
Radio/Fluoro modes: ± 2 % or ± 1 kV, whichever is greater |
|
Mammo modes: ± 2 % or ± 0.7 kV, whichever is greater |
| Reproducibility |
± 1 % (std of 5 readings) |
| Filtration correction range |
Radio/Fluoro modes: 1 mm Al to 10 mm Al or equivalent |
|
Mammo mode: 0 mm Al to 0.4 mm Al added filtration |
| | Dose/Exposure |
| Units |
Roentgens, Grays |
| Range |
0.5 mR to 999 R |
|
5 µGy to 999 Gy |
| Resolution |
1 µR |
|
0.01 µGy |
| Accuracy |
± 5 % |
| Reproducibility |
± 0.5 % (std of 5 readings) |
| Filtration correction range |
Radio/Fluoro modes: 1 mm Al to 10 mm Al or equivalent |
|
Mammo modes: 0 mm Al to 0.4 mm Al added filtration |
| kV correction range |
Radio/Fluoro modes: 40 kV to 150 kV |
|
Mammo modes: Mo/Mo: 22 kV to 35 kV |
| | Dose/Exposure rate |
| Units |
Roentgens or Grays per hour, minute, second, pulse |
| Range |
8 mR/s to 10 R/s |
|
70 µGy to 100 mGy/s |
|
130 µR/pulse to 160 mR/pulse (@ 60 PPS) |
|
1.2 µGy/pulse to 1.4 mGy/pulse (@ 60 PPS) |
| Resolution |
1 µR/s |
|
0.01 µGy/s |
|
0.02 µR/pulse (@ 60 PPS) |
|
0.2 n Gy/pulse (@ 60 PPS) |
| Accuracy |
± 5 % |
| Filtration Correction Range |
Radio/Fluoro Modes: 1 mm Al to 10 mm Al or equivalent |
|
Mammo Mode: 0 mm Al to 0.4 mm Al added filtration |
| kV Correction Range |
Radio/Fluoro Modes: 40 kV to 150 kV |
|
Mammo Mode: Mo/Mo: 22 kV to 35 kV |
| | Exposure time–radiographic modes |
| Units |
Milliseconds, pulses |
| Range Range (@ stated accuracy) |
Milliseconds: 10 ms to 9999 ms |
|
Pulses: 1 pulse to 999 pulses |
| Resolution |
Milliseconds: 0.1 ms |
|
Pulses: 1 pulse |
| Accuracy |
Milliseconds: 1 % or 0.5 ms |
|
Pulses: ± 1 pulse |
| Reproducibility |
Milliseconds: 1 % or 0.5 ms |
|
Pulses: ± 1 pulse |
| | Elapsed time–fluoro modes |
| Range |
10 sec to 9999 sec |
| Resolution |
0.1 second |
| Accuracy |
1 % or 0.5 sec |
| | Average pulse rate – pulsed fluoro |
| Range |
1 pps to 999 pps (pulses per second) |
| Resolution |
1 pps |
| Accuracy |
1 pps |
| | Average pulse width – pulsed fluoro |
| Range |
10 ms to 999 ms |
| Resolution |
0.1 ms |
| Accuracy |
1 % or 0.5 ms |
| | HVL |
| Range |
Radio/Fluoro modes: 1.2 mm Al to 10 mm Al (equivalent) |
|
Mammo mode: 0.2 mm Al to 0.6 mm Al (equivalent) |
| Resolution |
Radio/Fluoro modes: 0.1 mm Al (equivalent) |
|
Mammo mode: 0.01 mm Al (equivalent) |
| Accuracy |
Radio/Fluoro modes: ± 10 % or 0.2 mm Al (equivalent) |
|
Mammo mode: ± 5 % or 0.05 mm Al (equivalent) |
|
TNT 12000 DoseMate dosimeter specifications
| Accuracy |
| Exposure and exposure rate accuracy |
|
± 1 % of reading ± 2 range resolution steps over range of 18 °C to 28 °C and ± 2 % of reading ± 2 range resolution steps over the full operating temperature range of 0 °C to 50 °C |
|
3 % NIST-traceable calibration is provided with each system and includes effects of 96035B, and 96020C |
|
| Measurement modes |
| Dose |
|
Autoranging across five decades of sensitive ranges |
|
Automatic drift and offset compensation |
|
Automatic post-exposure display hold |
| | Rate |
|
Measurement range covers a span from low-level image intensifier measurements to unattenuated, direct beams |
|
Automatic offset compensation and nonlinear filtering |
|
Autoranging provides five decades of sensitivity ranges |
|
Display updates once per second |
|
| Low Rate Mode |
|
|
This mode is only for making very low dose rate measurements. Nonlinear digital filtering and autoranging provide five decades of sensitivity ranges. Display updates once per second. In this mode, automatic current offset and drift compensation are disabled. As a result, the system can display very low dose rates |
|
| Power requirements |
| Bias voltage supply |
|
Fixed electronic bias (~300 V) |
|
Bias voltage removed from triaxial input connector at instrument turnoff |
|
| Connections |
| Ion chamber input |
|
Triax, BNC |
|
Collector and guard positive-biased relative to ion chamber body and service dosimeter chassis |
| | Power | |
| General information |
| Display |
|
Wireless ZigBee or optional USB interface to TNT 12000 display indicates all ion chamber identification information, numerical measurement results, battery level, calibration date, and other information |
| | Weight | | | Dimensions (maximum) (WxDxH) |
|
15.2 cm x 11.4 cm x 4.45 cm |
|
(6 in x 4.5 in x 1.75 in) |
|
96020C and 96035B Diagnostic Ion Chambers
| Technical specifications |
| Energy range |
| 96020C |
30 kVp to 150 kVp |
| 96035B |
30 kVp to 150 kVp for diagnostic measurements |
|
20 kVp to 50 kVp for mammographic measurements |
| | Nominal volume |
| 96020C |
150 cm3; 11.30 cm diameter by 1.5 cm thick active volume |
| 96035B |
15 cm3; 3.96 cm diameter by 1.22 cm thick active volume |
| | Nominal sensitivity |
| 96020C |
H60: 2.08 R/C x 107 R/C at 22 °C and 760 mmHg (optimized for low-level image intensifier and cine measurements) |
| 96035B |
L100: 2.0 R/C x 108 R/C at 22 °C and 760 mmHg |
| | MV30 (PTB mammo point) |
| 96020C/96035B |
2.21 R/C x 108 R/C at 22 °C and 760 mmHg (flat energy response suitable for conventional diagnostic radiography and mammography) |
| | Leakage current |
| 96020C/96035B |
< 10 fA under normal bias conditions (300 V) |
| | Collection efficiency |
| 96020C |
95 % at 2,000 R/min |
| 96035B |
95 % at 5,000 R/min |
| | Wall material |
| 96020C |
Composite graphite-filled thermoplastic |
| 96035B |
Graphite-coated acrylic (methylmethacrylate) |
| | Window material |
| 96020C |
0.76 mm thick, graphite-coated polycarbonate |
| 96035B |
Both entrance windows are made of 0.25 mm graphite-coated polycarbonate |
| | Window density |
| 96020C |
91 mg/cm2 |
| 96035B |
32 mg/cm2 |
| | Active window area |
| 96020C |
100 cm2, centered within the chamber body |
| 96035B |
Each side of the chamber has a circular active window region centered 7.1 mm further from the BNC connector than the center of the chamber body; active window regions have an area of 12.32 cm2 |
| | Collector plate |
| 96020C |
0.8 mm thick graphite-coated acrylic plate, 10.80 cm in diameter; 2.16 cm x 2.85 cm guard region electrically isolated from collector area |
| 96035B |
0.25 mm thick, centrally mounted, graphite-coated, polycarbonate plate, 3.18 cm, ± 0.01 cm in diameter; 1.27 cm x 0.89 cm guard region is electrically isolated from the collector area |
| | Connector |
| 96020C/96035B |
Side-mounted, triaxial, two-lug BNC connector |
| | Calibration |
| 96020C |
Standard calibration performed at H60 (NIST defined as 60 kVp, first HVL of 6 mm Al, homogeneity coefficient of 94) |
| 96035B |
Standard calibration performed at one diagnostic and one mammographic beam quality |
|
Calibration factors normalized to 22 °C and 760 mmHg Diagnostic Unattenuated Beam |
|
Calibration on diagnostic side of chamber is performed at M80 (NIST defined as 80 kVp, first HVL of 2.97 mm Al, homogeneity coefficient of 57) Mammographic Beam |
|
Calibration on mammographic side performed at Mo/Mo28 (NIST defined as 28 kVp, first HVL of 0.332 mm Al, homogeneity coefficient of 74.3) or MV30 (PTB defined as 30 kVp, first HVL of 0.337 mm Al) |
|
mA/mAs specifications
| Measurement range |
| Invasive mAs (with shunt) mA/mAs ranges* |
| Min to Max |
0.00 to 99.99 |
|
100.0 to 999.9 |
|
1000 to 1999 |
|
*mAs measurements are fully auto-ranging, so values are display ranges only |
| | Non-invasive mAs (with clamp) mA/mAs ranges |
| Min to Max |
0.0 to 999.9 |
|
1000 to 3999 |
|
| Accuracy |
| Invasive mAs (with shunt) | |
|
| Model name | Description | Price* | | TNT 12000WD Wireless Detector + Display | TNT 12000WD Wireless Detector + TNT 12000D Display (Item # 3335763)
Includes:
- TNT 12000WD Wireless Detector (Item # 3345931)
- TNT 12000D Wireless Display (Item # 3345954)
- AC Adapter(s)/Charger(s) (Item # 3346048)
- Cable, Type A to Mini B USB (Item # 3346027)
- Cable, Mini A to Mini B USB (Item # 3346030)
- ZigBee USB Dongle (Item # 3341333)
- Customized Carrying Case (Item # 3346011)
- CD-ROM, DoseMate/TNT/mAs (Item # 3586667)
- Quick Start Reference Guide (Item # 3557133)
|
|  | | TNT 12000WD Wireless Detector | TNT 12000WD Wireless Detector (Item # 3335774)
Includes:
- TNT 12000WD Wireless Detector (Item # 3345931)
- AC Adapter(s)/Charger(s) (Item # 3346048)
- Cable, Type A to Mini B USB (Item # 3346027)
- ZigBee USB Dongle (Item # 3341333)
- Customized Carrying Case (Item # 3346011)
- CD-ROM, DoseMate/TNT/mAs (Item # 3586667)
- Quick Start Reference Guide (Item # 3557133)
|
|  | | TNT 12000D Wireless Display | TNT 12000D Wireless Display (Item # 3335795)
Includes:
- TNT 12000D Wireless Display (Item # 3345954)
- AC Adapter(s)/Charger(s) (Item # 3346048)
- Cable, Type A to Mini B USB (Item # 3346027)
- Cable, Mini A to Mini B USB (Item # 3346030)
- ZigBee USB Dongle (Item # 3341333)
- CD-ROM, DoseMate/TNT/mAs (Item # 3586667)
|
|  | | TNT 12000WD Wireless Detector with mAs + DoseMate + Display | TNT 12000WD Wireless Detector with mAs + TNT 12000 DoseMate + TNT 12000D Display (Item #3586807)
Includes:
- TNT 12000WD Wireless Detector with mAs (Item # 3586570)
- TNT 12000 DoseMate Dosimeter (Item # 3561890)
- TNT 12000D Wireless Display (Item # 3345954)
- AC Power Adapter (Item # 3548014)
- Cable, Type A to Mini B USB (Item # 3346027)
- Cable, Mini A to Mini B USB (Item # 3346030)
- Test Stand (Item # 3586537)
- Triax Cable, Male to Male, BNC, 6 ft (Item # 3586644)
- ZigBee USB Dongle (Item # 3341333)
- HVL Filter Set (Item # 3264115)
- Multi-Prong Adapter (Item # 3549414)
- Adapter Stem (Item # 3264091)
- Test Lead Kit, 1 Set of Test Leads (Item # 3555490 )
- mA Cable Assembly, BNC, male to female (Item # 1918780 )
|
|  | | TNT 12000WD Wireless Detector + DoseMate with mAs + Display | TNT 12000WD Wireless Detector + TNT 12000 DoseMate with mAs + TNT 12000D Display (Item #3586829)
Includes:
- TNT 12000WD Wireless Detector (Item # 3345931)
- TNT 12000 DoseMate Dosimeter with mAs (Item # 3586543)
- TNT 12000D Wireless Display (Item # 3345954)
- AC Power Adapter (Item # 3548014)
- Cable, Type A to Mini B USB (Item # 3346027)
- Cable, Mini A to Mini B USB (Item # 3346030)
- Test Stand (Item # 3586537)
- Triax Cable, Male to Male, BNC, 6 ft (Item #3586644)
- ZigBee USB Dongle (Item # 3341333)
- HVL Filter Set (Item # 3264115)
- Multi-Prong Adapter (Item # 3549414)
- Adapter Stem (Item # 3264091)
- Test Lead Kit, 1 Set of Test Leads (Item # 3555490)
- mA Cable Assembly, BNC, male to female (Item # 1918780)
|
|  | | TNT 12000WD Wireless Detector with mAs + Display | TNT 12000WD Wireless Detector with mAs + TNT 12000D Display (Item # 3586787)
Includes:
- TNT 12000WD Wireless Detector with mAs (Item # 3586570)
- TNT 12000D Wireless Display (Item # 3345954)
- AC Power Adapter (Item # 3548014)
- Cable, Type A to Mini B USB (Item # 3346027)
- Cable, Mini A to Mini B USB (Item # 3346030)
- ZigBee USB Dongle (Item # 3341333)
- Multi-Prong Adapter (Item # 3549414)
- Test Lead Kit, 1 Set of Test Leads (Item # 3555490)
- mA Cable Assembly, BNC, male to female (Item # 1918780)
- mA Shunt Assembly (Item # 3586555)
- Connector, BNC (female) to Binding Posts (Item # 1938315)
- Connector, BNC (female) Double Banana Plug (Item # 1633042)
- Customized Carrying Case (Item # 3586865)
- CD-ROM, DoseMate/TNT/mAs (Item # 3586667)
|
|  | | TNT 12000 DoseMate with mAs + Display | TNT 12000 DoseMate with mAs + TNT 12000D Display (Item # 3586779)
Includes:
- TNT 12000 DoseMate Dosimeter with mAs (Item # 3586543)
- TNT 12000D Wireless Display (Item # 3345954)
- AC Power Adapter (Item # 3548014)
- Cable, Type A to Mini B USB (Item # 3346027)
- Cable, Mini A to Mini B USB (Item # 3346030)
- Test Stand (Item # 3586537)
- Plug, Connector, Triax (Item # 3551210)
- Triax Cable, Male toMale, BNC, 6 ft (Item # 3586644)
- ZigBee USB Dongle (Item # 3341333)
- HVL Filter Set (Item # 3264115)
- Multi-Prong Adapter (Item # 3549414)
- Adapter Stem (Item # 3264091)
- Test Lead Kit, 1 Set of Test Leads (Item # 3555490)
- mA Cable Assembly, BNC, male to female (Item # 1918780)
- mA Shunt Assembly (Item # 3586555)
|
|  | | TNT 12000WD Wireless Detector + DoseMate + Display | TNT 12000WD Wireless Detector + TNT 12000 DoseMate + TNT 12000D Display (Item # 3586841)
Includes:
- TNT 12000WD Wireless Detector (Item # 3345931)
- TNT 12000 DoseMate Dosimeter (Item # 3561890)
- TNT 12000D Wireless Display (Item # 3345954)
- AC Power Adapter (Item # 3548014)
- Cable, Type A to Mini B USB (Item # 3346027)
- Cable, Mini A to Mini B USB (Item # 3346030)
- Test Stand (Item # 3586537)
- Triax Cable, Male to Male, BNC, 6 ft (Item # 3586644)
- ZigBee USB Dongle (Item # 3341333)
- HVL Filter Set (Item # 3264115)
- Multi-Prong Adapter (Item # 3549414)
- Adapter Stem (Item # 3264091)
- Customized Carrying Case (Item # 3586528)
- CD-ROM, DoseMate/TNT/mAs (Item # 3586667)
- Quick Start Reference Guide (Item # 3557133)
|
|  | | TNT 12000 DoseMate + Display | TNT 12000 DoseMate + TNT 12000D Display (Item # 3586754)
Includes:
- TNT 12000 DoseMate Dosimeter (Item # 3561890)
- TNT 12000D Wireless Display (Item # 3345954)
- AC Power Adapter (Item # 3548014)
- Cable, Type A to Mini B USB (Item # 3346027)
- Cable, Mini A to Mini B USB (Item # 3346030)
- Test Stand (Item # 3586537)
- Triax Cable, Male toMale, BNC, 6 ft (Item # 3586644)
- ZigBee USB Dongle (Item # 3341333)
- HVL Filter Set (Item # 3264115)
- Multi-Prong Adapter (Item # 3549414)
- Adapter Stem (Item # 3264091)
- Customized Carrying Case (Item # 3586528)
- CD-ROM, DoseMate/TNT/mAs (Item # 3586667)
|
|  | | TNT 12000WD Wireless Detector with mAs + DoseMate | TNT 12000WD Wireless Detector with mAs + TNT 12000 DoseMate (Item # 3586818)
Includes:
- TNT 12000WD Wireless Detector with mAs (Item # 3586570)
- TNT 12000 DoseMate Dosimeter (Item # 3561890)
- AC Power Adapter (Item # 3548014)
- Cable, Type A to Mini B USB (Item # 3346027)
- Cable, Mini A to Mini B USB (Item # 3346030)
- Test Stand (Item # 3586537)
- Triax Cable, Male toMale, BNC, 6 ft (Item # 3586644)
- ZigBee USB Dongle (Item # 3341333)
- HVL Filter Set (Item # 3264115)
- Multi-Prong Adapter (Item # 3549414)
- Adapter Stem (Item # 3264091)
- Test Lead Kit, 1 Set of Test Leads (Item # 3555490)
- mA Cable Assembly, BNC, male to female (Item # 1918780)
- mA Shunt Assembly (Item # 3586555)
- Connector, BNC (female) to Binding Posts (Item # 1938315)
|
|  | | TNT 12000WD Wireless Detector + DoseMate with mAs | TNT 12000WD Wireless Detector + TNT 12000 DoseMate with mAs (Item # 3586852)
Includes:
- TNT 12000WD Wireless Detector (Item # 3345931)
- TNT 12000 DoseMate Dosimeter with mAs (Item # 3586543)
- AC Power Adapter (Item # 3548014)
- Cable, Type A to Mini B USB (Item # 3346027)
- Cable, Mini A to Mini B USB (Item # 3346030)
- Test Stand (Item # 3586537)
- Triax Cable, Male to Male, BNC, 6 ft (Item # 3586644)
- ZigBee USB Dongle (Item # 3341333)
- HVL Filter Set (Item # 3264115)
- Multi-Prong Adapter (Item # 3549414)
- Adapter Stem (Item # 3264091)
- Test Lead Kit, 1 Set of Test Leads (Item # 3555490)
- mA Cable Assembly, BNC, male to female (Item # 1918780)
- mA Shunt Assembly (Item # 3586555)
- Connector, BNC (female) to Binding Posts (Item # 1938315)
- Connector, BNC (female) Double Banana Plug (Item # 1633042)
- Customized Carrying Case (Item # 3586528)
- CD-ROM, DoseMate/TNT/mAs (Item # 3586667)
- Quick Start Reference Guide (Item # 3557133)
| |  | | TNT 12000WD Wireless Detector + DoseMate | TNT 12000WD Wireless Detector + TNT 12000 DoseMate (Item # 3586834)
Includes:
- TNT 12000WD Wireless Detector (Item # 3345931)
- TNT 12000 DoseMate Dosimeter (Item # 3561890)
- AC Power Adapter (Item # 3548014)
- Cable, Type A to Mini B USB (Item # 3346027)
- Cable, Mini A to Mini B USB (Item # 3346030)
- Test Stand (Item # 3586537)
- Triax Cable, Male toMale, BNC, 6 ft (Item # 3586644)
- ZigBee USB Dongle (Item # 3341333)
- HVL Filter Set (Item # 3264115)
- Multi-Prong Adapter (Item # 3549414)
- Adapter Stem (Item # 3264091)
- Customized Carrying Case (Item # 3586528)
- CD-ROM, DoseMate/TNT/mAs (Item # 3586667)
- Quick Start Reference Guide (Item # 3557133)
|
|  | | TNT 12000WD Wireless Detector with mAs | TNT 12000WD Wireless Detector with mAs (Item # 3586793)
Includes:
- TNT 12000WD Wireless Detector with mAs (Item # 3586570)
- AC Power Adapter (Item # 3548014)
- Cable, Type A to Mini B USB (Item # 3346027)
- Cable, Mini A to Mini B USB (Item # 3346030)
- ZigBee USB Dongle (Item # 3341333)
- Multi-Prong Adapter (Item # 3549414)
- Test Lead Kit, 1 Set of Test Leads (Item # 3555490)
- mA Cable Assembly, BNC, male to female (Item # 1918780)
- mA Shunt Assembly (Item # 3586555)
- Connector, BNC (female) to Binding Posts (Item # 1938315)
- Connector, BNC (female) Double Banana Plug (Item # 1633042)
- Customized Carrying Case (Item # 3586865)
- CD-ROM, DoseMate/TNT/mAs (Item # 3586667)
- Quick Start Reference Guide (Item # 3557133)
|
|  | | TNT 12000 DoseMate with mAs | TNT 12000 DoseMate with mAs (Item # 3586768)
Includes:
- TNT 12000 DoseMate Dosimeter with mAs (Item # 3586543)
- AC Power Adapter (Item # 3548014)
- Cable, Type A to Mini B USB (Item # 3346027)
- Cable, Mini A to Mini B USB (Item # 3346030)
- Test Stand (Item # 3586537)
- Triax Cable, Male toMale, BNC, 6 ft (Item # 3586644)
- ZigBee USB Dongle (Item # 3341333)
- HVL Filter Set (Item # 3264115)
- Multi-Prong Adapter (Item # 3549414)
- Adapter Stem (Item # 3264091)
- Test Lead Kit, 1 Set of Test Leads (Item # 3555490)
- mA Cable Assembly, BNC, male to female (Item # 1918780)
- mA Shunt Assembly (Item # 3586555)
- Connector, BNC (female) to Binding Posts (Item # 1938315)
- Connector, BNC (female) Double Banana Plug (Item # 1633042)
|
|  | | TNT 12000 DoseMate Dosimeter | TNT 12000 DoseMate Dosimeter (Item # 3588752)
Includes:
- TNT 12000 DoseMate Dosimeter (Item # 3561890)
- AC Power Adapter (Item # 3548014)
- Cable, Type A to Mini B USB (Item # 3346027)
- Cable, Mini A to Mini B USB (Item # 3346030)
- Test Stand (Item # 3586537)
- Triax Cable, Male toMale, BNC, 6 ft (Item # 3586644)
- ZigBee USB Dongle (Item # 3341333)
- HVL Filter Set (Item # 3264115)
- Multi-Prong Adapter (Item # 3549414)
- Adapter Stem (Item # 3264091)
- Customized Carrying Case (Item # 3586528)
- CD-ROM, DoseMate/TNT/mAs (Item # 3586667)
|
|  |
| Accessories | Price* | | 500-100 | CT Ion Chamber 3.2 cc |  | | 500-200 | CT Ion Chamber 10 cc |  | | 86020 | 86020 Triax Cable |  | | 9602DC | Diagnostic Ionization Chamber 150cc |  | | 96035B | Diagnostic Ionization Chamber 15 cc |  | | EM4524-C-240 | EM4524-C-240 mA Cable Assembly |  |
|
|
|
|
|
|
|