VT PLUS HF Gas Flow Analyzer

A benchtop ventilator tester with bi-directional flow, pressure, volume, oxygen concentration, and pressure measurements

VT PLUS HF Gas Flow Analyzer

VT PLUS HF Gas Flow Analyzer

The VT PLUS HF is Fluke Biomedical’s premier general-purpose gas-flow analyzer. In addition, special display modes and bi-directional flow make it perfect for fully and efficiently testing both conventional mechanical ventilators and high-frequency ventilators. EC.6.20 now requires 100 % completion of scheduled life-support device preventive maintenance every year, and VT PLUS HF can help meet those requirements. Multiple special-function tests make troubleshooting quick and efficient.

VT PLUS HF has the capability to measure either high- or low-flow and pressure, replacing the need for gauges and flow meters. It measures 21 ventilator parameters and can display all of them on one screen. Results can be printed directly from the unit or from a PC with included Windows-compatible software. VT PLUS HF also has onboard graphing capability and shows the minimum, maximum, average, and absolute measurement for all parameters.

Learning to use the VT PLUS HF is simple. Technicians control the unit using the VT PLUS HF user-friendly command system, or, if they’re familiar with the RT-200, they can switch to a special control mode that uses RT-200-style commands.

VT PLUS HF can be operated with a variety of precision simulation lungs to ensure that ventilators are tested to manufacturers’ specifications and clinical expectations with a fully NIST-traceable testing system.

Key Features

  • Bi-directional flow, pressure, volume, and oxygen concentration, and pressure measurements
  • Low- and high-pressure, and flow measurement capability
  • Special HF mode—up to 900 BPM (15 Hz)
  • RS-232 and printer ports
  • Included Windows-compatible graphics software
  • All 21ventilator parameters displayed at once on one screen
  • Operation by user-friendly VT PLUS HF command mode or special RT-200 command mode
  • Minimum, maximum, average, absolute, and graph for all parameters
  • Multiple special-function tests for efficient troubleshooting

Optional Features

  • Operation with a variety of precision simulation lungs available from Fluke Biomedical to complete a fully NIST-traceable ventilator testing system

 

Specifications
Power
100 V ac to 240 V ac, 50/60 Hz
Maximum Over-Voltage: 264 Vac
Display
320 x 240 LCD with CFL backlight
Viewing Area: 10.1 cm x 8.2 cm (3 in x 4 in)
Blue on white background
Operational Modes
Manual mode, for simple tests or troubleshooting
Computer-control mode, using RS-232 serial port for special applications
Use of VT PLUS HF with VT Plus for Windows, for recording graphs and logging data to a computer
Oxygen Measurement
Range: 0 % to 100 %
Accuracy: ±2 % FSO
Resolution: 0.1 % O2
Transducer Location: Internal
Gas Compatibility
Air, O2, CO2, N2, N2O, He, mixtures or user-defined
Gas Reference Units
ATP, STPD0, STPD21 and BTPS
Output Ports
RS-232 serial port and parallel printer port
Test Parameters
Continuous Flow: Low Flow: ± (2 % of reading and 1 % of range)
High Flow: ± (2 % of reading and 1 % of range)
Volumetric Flow:
Low-Flow:
Flow Range: -25 lpm to 25 lmp
Accuracy: ± 1 % of range
Frequency Response: > 25 Hz or t10-90 < 40 ms, whichever is greater
Low Flow Dropout: 0.01 lpm
Breath-Defect Threshold: 0.5 lpm
Maximum-Flow Rate: 50 lpm
Volume Range: > ± 60 L
Sample Rate: 100 Hz
Resolution: 0.01 lpm flow > 1 lpm; 0.001 lpm flow < 1 lpm
Dynamic resistance: < 2.5 cmH2O @ 5 lpm
Fittings: 15 mm OD, 1:40 conical male; 0.25 in NPT ID per ASTM F-1054 aluminum with black anodized finish
Notes:
  • Tidal-volume accuracy: ± 3 % of reading or ± 2 ml, whichever is greater
  • Volume accuracy tested to 1 liter
  • Flow accuracy is specified for dry air or oxygen
  • Below 3.0 lpm, measurement accuracy is obtained by allowing the VT PLUS HF to fully warm up or manually zeroing before reading or documenting measurement
High-Flow:
Flow Range: -300 lpm to 300 lmp
Accuracy: ± 2 % of reading or ± 2 % of range, whichever is greater
Frequency Response: > 25 Hz
High-Flow Dropout: 25 lpm
Breath-Detect Threshold: 2 lpm
Maximum-Flow Rate: 500 lpm
Volume Range: > ± 60 L
Dynamic Resistance: < 2 cmH2O @ 60 lpm
Sample Rate: 100 Hz
Resolution: 0.01 lpm
Fittings: 22 mm OD, 1:40 conical male; 15 mm ID, 1:40 conical female per ASTM F-1054 aluminum with black anodized finish
Notes:
  • Tidal-volume accuracy: ± 3 % of reading or ± 10 ml, whichever is greater
  • Volume accuracy tested to 7 liters
  • Flow accuracy is specified for dry air or oxygen
Low-Pressure:
Range: ± 500 mmHg (10 psi)
Accuracy: ± 0.8 % of reading or ± 1.5 mmHg, whichever is greater
Frequency Response: > 10 Hz
Resolution: 0.1 mmHg
Fittings: Luer lock, stainless steel
Maximum Applied Pressure: 60 psi
Sample Rate: 100 Hz
Operating Pressure: 30 psi
Note:
Fluid pressure may be applied to the positive port; however, fluids should be kept from entering the pressure port by using a suitable length of connection tubing

High-Pressure:
Maximum Applied Pressure: 150 psi
Range: ± 100 psi
Accuracy: ± 1 % of reading or ± 0.3 psig, whichever is greater
Frequency Response: > 10 Hz
Resolution: 0.1 psi
Sample Rate: 100 Hz
Fittings: DISS connector, stainless steel

Airway-Pressure:
Maximum Applied Pressure: 20 psi
Range: ± 120 cmH2O
Accuracy: ± 0.75 % of reading or ± 0.5 cmH2O, whichever is greater
Frequency Response: > 25 Hz or t10-90 < 40 ms, whichever is greater
Resolution: 0.1 cmH2O
Fittings: Internally connected at the transducer distal end
Note:
Airway pressure is internally tapped off the proximal-flow sensor port, which is the port closest to the exhaust port on the VT PLUS HF
Ventilator Parameter
Inspiratory and Expiratory Tidal Volume: Range:As specified in high-flow/low-flow spec
Accuracy: As specified in high-flow/low-flow spec
Expiratory Minute Volume Range: 0 L to 60 L
Accuracy: ± 3 of % or 250 ml
Breath Rate Range: 0.5 BPM to 150.0 BPM
Accuracy:± 1 of %
Inspiratory-to-Expiratory Time Ratio (I:E Ratio): Range: 1:200 to 200:1
Accuracy: ± 2 % or ± 0.1 s
Inspiratory Time: Range: 0 s to 60 s
Accuracy: ± 1 % or ± 0.1 s
Expiratory Time: Range: 0 s to 90 s
Accuracy: ± 1 % or ± 0.1 s
Peak Inspiratory
Pressure:
Range: ± 120 cmH2O
Accuracy: ± 3 % or ± 1 cmH2O
Inspiratory Pause
Pressure:
Range: ± 120 cmH2O
Accuracy: ± 3 % or ± 1 cmH2O
Mean Airway Pressure: Range: ± 80 cmH2O
Accuracy: ± 3 % or ± 0.5 cmH2O
Positive End-Expiratory Pressure (PEEP): Range: -5 to 40 cmH2O
Accuracy: ± 3 % or ± 0.5 cmH2O
Lung Compliance: Range: 0 to 150 ml/cmH2O
Accuracy: ± 5 % or ± 5 ml/cmH2O
Inspiratory Hold Time: Range: 0 s to 60 s
Accuracy: ± 1 % or ± 0.1 s
Expiratory Hold Time: Range: 0 s to 90 s
Accuracy: ± 1 % or ± 0.1 s
Peak Expiratory Flow: Range: 0 lpm to 300 lpm
Accuracy: ± 3 % or ± 2 lpm
Peak Inspiratory Flow: Range: 0 lpm to 300 lpm
Accuracy: ± 3 % or ± 2 lpm
Flow Bias: Range: 0 lpm to 30 lpm
Accuracy: ± 2 % or ± 0.5 lpm
Operating Environment Conditions
Temperature Range: 10 °C to 40 °C
Ambient Humidity: 0 % to 80 % noncondensing to 31 °C, decreasing to 50 % at 40 °C
Barometric Pressure: 8 psig to 18 psig
Storage Environment Conditions
Temperature Range: -25 °C to -50 °C
Humidity: 0 to 95 % noncondensing
Dimensions (WxDxH)
25.4 cm x 25.4 cm x 12.7 cm
(10 in x 10 in x 5 in)
Weight
4.53 kg (10 lb)
Model NameDescription
VT+HF-US120VT PLUS HF Gas-Flow Analyzer – United States, 120 V

All models include the following:

  • Operator's manual
  • VT for Windows PC Software (8830200FG)
  • Serial cable (75034)
  • Tilt stand (1HD0011)
  • Power cord (country specific)
  • Accessory kit (to include the following 16 accessories) (VT-PLUS-7001)
  • Filter, external (bacterial) (one each)
  • Filter, external (60mm/low flow) (one each)
  • Filter, external (pulmonary function) (one each)
  • Adapter, DISS O2 nut and nipple with 1/4" I.D. hose barb (one each)
  • Tubing adapter, directional (15mm OD x 15mm OD) (two each)
  • Tubing adapter (22mm OD x 22mm ID) (two each)
  • Tubing adapter (22mm OD x 22mm OD) (two each)
  • Tubing adapter (15mm OD x 22mm OD) (two each)
  • Tubing adapter (15mm OD x 15mm OD) (two each)
  • Tubing adapter (15mm ID x 15mm OD) (two each)
  • Tubing adapter, narrow bore (two each)
  • Barb (luer lock – male to 1/89 ID tubing) (two each)
  • Tubing adapter (1/4 NPT male to 1/8 ID tubing barb fitting) (two each)
  • Tubing adapter (luer lock 1/16 to bulkhead connection) (two each)
  • Fuse (500 mA)
  • Tubing 1/8" 4' long (two each)
VT+HF/ACCULUNG-US VT PLUS HF Gas-Flow Analyzer/ACCU LUNG – United States

Includes:
  • Premium Precision Ventilator Test Kit: VT PLUS HF Gas-Flow Analyzer.
  • AccuLung portable precision lung/soft-sided carrying case for AccuLung.
VT+HF-AUS250V VT PLUS HF Gas-Flow Analyzer – Australia, 250 V

Includes:
  • VT PLUS HF Gas-Flow Analyzer only
VT+HF/ACCULUNG-AUS VT PLUS HF Gas-Flow Analyzer/ACCU LUNG – Australia

Includes:
  • Premium Precision Ventilator Test Kit: VT PLUS HF Gas-Flow Analyzer.
  • AccuLung portable precision lung/soft-sided carrying case for AccuLung.
VT+HF-SHK250V VT PLUS HF Gas-Flow Analyzer – Schuko, 250 V

Includes:
  • VT PLUS HF Gas-Flow Analyzer only
VT+HF/ACCULUNG-SHK VT PLUS HF Gas-Flow Analyzer/ACCU LUNG – Schuko

Includes:
  • Premium Precision Ventilator Test Kit: VT PLUS HF Gas-Flow Analyzer.
  • AccuLung portable precision lung/soft-sided carrying case for AccuLung.
VT+HF-UK250V VT PLUS HF Gas-Flow Analyzer – United Kingdom, 250 V

Includes:
  • VT PLUS HF Gas-Flow Analyzer only
VT+HF/ACCULUNG-UK VT PLUS HF Gas-Flow Analyzer/ACCU LUNG – United Kingdom

Includes:
  • Premium Precision Ventilator Test Kit: VT PLUS HF Gas-Flow Analyzer.
  • AccuLung portable precision lung/soft-sided carrying case for AccuLung.
VT+HF-BRA250V VT PLUS HF Gas-Flow Analyzer – Brazil, 250 V

Includes:
  • VT PLUS HF Gas-Flow Analyzer only
VT+HF/ACCULUNG-BRAZ VT PLUS HF Gas-Flow Analyzer/ACCU LUNG – Brazil

Includes:
  • Premium Precision Ventilator Test Kit: VT PLUS HF Gas-Flow Analyzer.
  • AccuLung portable precision lung/soft-sided carrying case for AccuLung.
8831007VT-Plus upgrades (adds FH capability and RT-200 mode)

VT PLUS HF hardware and firmware factory service upgrade (for units lower than hardware v1.01.01; additional flat-rate charge required for factory service/calibration)

Gas Flow Analyzers
ACCULUNG IIAccuLung II Adjustable Portable Precision Test Lung
(part# 4281291)
ACCU LUNGAccuLung Portable Precision Test Lung (part# 2387318)
with soft-sided carrying case
48499Siemens 190 portable precision lung (part# 2213774)

Cables
33705123370512 Dual 9DF serial cable (part# 2204472)
Special cable to attach to RS232 on a PC.
71072Parallel printer cable, D25M to C36M (part# 2238072)

Carrying Cases
9530-0066Multi-purpose Hard-sided Watertight Carrying Case
(part# 2248587)
(contains pick-and-pluck foam) (19 " L x 14 " W x 7 ¾ " H)
5022010Soft vinyl carrying case (part# 2222822)

Miscellaneous Accessories
MI-12952Michigan Instruments Non-Instrumented Adult/Infant AccuLung portable precision lung (part# 2251013)
MI-14900Michigan Instruments Non-Instrumented Single-Adult AccuLung portable precision lung (part# 2251049)
MI-11000Michigan Instruments Non-Instrumented Dual-Adult AccuLung portable precision lung
RP RING KITParabolic Airway Resistor Ring kit (part# 3799622)

Printers
60196120V Power Supply for Seiko DPU414 printer (part# 2235375)
Printer used with the following products:
•VT PLUS HF Gas Flow Analyzer
•ISA601 Pro SeriesXL
•IDA 4 Plus
•Index 2XL
•QED 6
•BP Pump 2
PRINTR/CTZ-US120VPrinter, 115V, Citizen IDP3110 (part# 2248762)
PRINTR/CTZ-US220VPrinter 220V, Citizen IDP3110 Schuko (part# 2719653)
97116Printer paper DPU-414 and DPU-411 (part# 2248737)
(minimum 7 rolls – price is per roll)
PRINTR/414-US120VPrinter, SEIKO DPU-414-30B (part# 2248899)
Includes 120V PWR Supply (part# 2235375)
(Note: requires additional purchase of parallel printer cable, part# 2238072)
for use with ESA601/IDA 4 Plus/Index 2XL/QED 6/BP Pump 2
PRINTR414-SHK220VPrinter, Seiko DPU-414-30B with 220 V power supply (part# 2399531)
(NOTE: requires additional purchase of serial printer cable (part# 3370513)

Software
ANSUR VTAnsur VT Plug-In license (part# 3837485)

Service Packages
Care PlansRequest a quote
Biomedical Equipment  |  Patient Simulator  |  Safety Analyzer  |  Dosimeter  |  Radiation Detector
Related Products
View related products
 
 Ansur VT Plug-In license »
 
 VT MOBILE Gas Flow Analyzer »
 
 VT305 Gas Flow Analyzer »
MDQA Webinar
Follow us Fluke Biomedical on Facebook Fluke Biomedical on Twitter Fluke Biomedical on YouTube Fluke Biomedical on LinkedIn Home | Site Map | Terms of Use | Terms and Conditions | Privacy Policy | Disclaimer © 1995 - 2018 Fluke Corporation