ESA615 Electrical Safety Analyzer

ESA615 Electrical Safety Analyzer
ESA615
ESA615 is portable
ESA615 is simple to use
ESA615 is automated

ESA615 Electrical Safety Analyzer

The ESA615 Electrical Safety Analyzer brings fast and simple automated testing in the form of a portable analyzer to healthcare technology management professionals that perform electrical safety testing on medical equipment both in the field and in facilities. Whether it is simple testing or comprehensive analysis, the ESA615 can do it all. This multifaceted device performs all primary electrical safety tests, including line (mains) voltage, ground wire (protective earth) resistance, insulation resistance, device current, and lead (patient) leakage tests. It also offers ECG simulation and point–to-point voltage, leakage, and resistance tests. Versatile to any global electrical safety standard of choice, the ESA615 can test to ANSI/AAMI ES-1, NFPA-99, IEC62353, VDE751, IEC60601-1 (2nd and 3rd editions) and AS/NZS 355 standards. The ESA615 is an all-in-one solution with a multimeter, safety analyzer and ECG simulator in a single electrical safety test instrument.

Automated: Pre-set templates allow you to test to the global electrical safety standard of your choices at the push of a button. Let the unit guide you through the test steps quickly and accurately in three simple steps:

  1. Select the test sequence corresponding to your standard of choice
  2. Initiate the automated test sequence
  3. Store the results on-board or wirelessly export to your PC in seconds

It is that easy. The automated test sequences are easily customized to suit individual testing requirements.

Portable: The ESA615 is a small, lightweight analyzer complete with an integrated handle to make transportation from place to place simple for on-the-go field service professionals. It is designed for operation in tight spaces, and can easily fit on any cart. A light, protective carrying case makes it easy to store and transport safely.

Simple: A push-button interface allows quick access to highly-comprehensive functions and features, and an intuitive interface guides the user through tests. The ESA615 features a large display for clear indication of available functions, set-up criteria, device under test receptacle conditions, and test results. Data can be entered quickly with a plug–and-play keyboard, barcode scanner and/or on-board data entry interface. Data archival is fast and simple with wireless communication or through a removable memory card with capacity for up to 100 automated test sequences and 1000 test results.

  • On-board automation with automated test sequences for easy compliance to key global electrical safety standards (ANSI/AAMI ES-1 (NFPA-99), IEC62353 (VDE751), IEC60601-1 2nd and 3rd editions, and AS/NZS 3551)
  • Portable, ergonomic design with an integrated handle and tilt stand
  • Human-factors-designed user interface for streamlined testing
  • ECG waveform tests and dual-lead measurements combine the functionality of a simulator, multimeter and safety analyzer in a single test tool
  • 20 A at 120 V current capability
  • Five applied parts jacks and easy ECG snap connection; optional expander box for up to 12-lead ECG testing
  • Easy data entry through barcode, external keyboard or on-board keypad
  • Wireless communication plus removable memory card for fast and convenient data storage and exchange
  • Replaceable mains fuses keep your unit in the field and out of the repair shop
  • Custom language selections include: English, French, German, Spanish, Italian and Portuguese
  • Large, easy-to-read display with adjustable contrast
  • Optional Ansur automation software completely integrates medical device testing including electrical safety, visual inspection and other performance parameters for total digital data management.  Tethers with other Ansur-automated test devices for harmonized workflow and reporting. 
  • Rigorously tested for rugged field applications, with CE and CSA in addition to Fluke-quality- ruggedness for long-term reliability
  • Two-year extended warranty (no-cost, available after first-year calibration at any authorized Fluke Biomedical Service Center)
  • Global support network delivering prompt service and peace of mind to Fluke Biomedical customers worldwide

Voltage 

Range (mains voltage)

90 V ac rms to 132 V ac rms

180 V ac rms to 264 V ac rms

Range (accessible voltage)

0 V ac rms to 300 V ac rms

Accuracy

± (2 % of reading + 0.2 V)

Voltage tests

Mains and point to point

Earth resistance

Modes

Two-Wire

Test current / Ranges / Accuracy

> 200 mA ac

0 Ω to 2 Ω

± (2 % of reading + 0.015 Ω)

Resistance tests

Earth resistance and point to point

Equipment current

Mode

AC rms

Range / Accuracy

0 A to 20 A

± (5 % of reading + (2 counts or 0.2 A, whichever is greater))

Duty cycle

15 A to 20 A, 5 min. on/5 min. off

10 A to 15 A, 7 min. on/3 min. off

0 A to 10 A continuous

Leakage current

Modes*

AC + DC (True rms)

AC only

DC only

* Modes are available in all leakage tests with the exception of MAP leakages that are available only in true rms

Patient load selection (input impedance)

AAMI ES1-1993 Fig.1

IEC 60601: Fig 15

Crest factor

≤ 3

Ranges

0 µA to 199.9 μA

200 µA to 1999 μA

2 mA to 10 mA

Frequency response / accuracy

DC to 1 kHz

± (1 % of reading + (1 μA or 1 LSB, whichever is greater))

1 kHz to 100 kHz

± (2 % of reading + (1 μA or 1 LSB, whichever is greater))

1 kHz to 5 kHz (current > 1.6 mA)

± (4 % of reading + (1 µA or 1 LSD, whichever is greater))

100 kHz to 1 MHz

± (5 % of reading + (1 μA or 1 LSB, whichever is greater))

Accuracy for Isolation, MAP, Direct AP, Alternative AP, and Alternative Equipment leakage tests all ranges are:

  • At 120 V ac + (2.5 µA or 1 LSD, whichever is greater)
  • At 230 V ac additional ±3.0 % and + (2.5 µA or 1 LSD, whichever is greater)

For Alternative equipment, Alternative AP, and Direct AP leakage tests, the leakage values are compensated for

nominal mains as per 62353. Therefore, the accuracy specified for other leakages is not applicable.



Leakage tests

Ground wire (earth)

Chassis (enclosure)

Lead to ground (patient)

Lead to lead (patient auxiliary)

Lead isolation (mains on applied part)

Direct equipment

Direct applied part

Alternative equipment

Alternative applied part

Point to point

Mains on applied part

Test voltage

100 % ± 7 % of Mains for AAMI, current limited to 1 mA ±25 % per AAMI

100 % ± 7 % of Mains for IEC 62353 current limited to 3.5 mA ± 25 % per IEC 62353

100 % ± 7 % of Mains for IEC 60601-1 current limited to 7.5 mA ± 25 % per IEC 60601-1

Differential leakage

Ranges

75 µ A to 199 µA

200 µA to 1999 µA

2 mA to 20 mA

Accuracy

± (10 % of reading + (2 counts or 20 µA, whichever is greater))

Insulation resistance

Ranges / Accuracy

0.5 MΩ to 20 MΩ

± (2 % of reading + 0.2 MΩ)

20 MΩ to 100 MΩ

± (7.5 % of reading + 0.2 MΩ)

Source test voltage

500 V dc  or 250 V dc

(+ 20 %, -0 %) 2.0 ± 0.25 mA short-circuit current

Insulation resistance tests

Mains-PE, AP-PE, Mains- PE, Mains-NE (non-earthed accessible conductive part) and AP- NE (non-earthed accessible conductive part)



ECG performance waveforms

Accuracy

± 2 %

± 5 % for amplitude of 2 Hz Square wave only, fixed @ 1 mV Lead II configuration

Waveforms

Rates

ECG Complex

30 BPM, 60 BPM,

120 BPM, 180 BPM, and 240 BPM

Ventricular Fibrillation

 

Square wave (50 % duty cycle)

0.125 Hz and 2 Hz

Sine wave

10 Hz, 40 Hz, 50 Hz, 60 Hz, and 100 Hz

Triangle wave

2 Hz

Pulse (63 ms pulse width)

30 BPM and 60 BPM

Test Standards

Available Selections

ANSI/AAMI ES-1, IEC62353, IEC60601-1, and AN/NZS 3551

Built-in Autosequences

IEC60601-1 2nd Edition

Patient Monitor, Defibrillator, Infusion Pump, Ultrasound Device, Generic Device and System

IEC62353

Patient Monitor, Defibrillator, Infusion Pump, Ultrasound Device and Generic Device

NFPA-99 (Hospital)

Patient Monitor, Defibrillator, Infusion Pump, Ultrasound Device and Generic Device

ANSI/AAMI ES-1

Patient Monitor, Defibrillator, Infusion Pump, Ultrasound Device and Generic Device

Communications

USB Device Upstream Port

Mini-B connector for control by a computer

USB Host Controller Port

Type A, 5 V output, 0.5 A max load. Connector for keyboard and barcode reader

Wireless

IEEE 802.15.4 for control by a computer

Modes of Operation

Manual and remote

Power ratings

Mains voltage outlet

120 V ac

230 V ac

Mains voltage inlet power range 

90 V ac rms to 132 V ac rms

180 V ac rms to 264 V ac rms

Maximum current

20 A

16 A

Hz

47 to 63 Hz,

47 to 63 Hz,

Physical case

Dimensions (WxDxH)

17.6 cm x 8.4 cm x 28.5 cm (6.9 in x 3.3 in x 11.2 in)

Weight

1.6 kg (3.5 lb)

Environmental

Operating temperature

10 ºC to 40 ºC (50 ºF to 104 ºF)

Storage temperature

-20 ºC to 60 ºC (-4 ºF to 140 ºF)

Operating humidity

10 % to 90 % non-condensing

Altitude

120 V ac mains supply voltage up to 5000 meters

230 V ac mains supply voltage up to 2000 meters

Warranty

Two-year extended warranty (no-cost, available after first-year calibration at any authorized Fluke Biomedical Service Center, otherwise standard one year warranty applies)

Model NameDescription
ESA615-11-BRAZESA615-11 Brazil 230V Includes:
  • Operator’s Manual (multilingual CD) (CD-ROM)
  • Getting Started Guide, hard copy, multilingual (Manual)
  • Ansur Plug-in, CD with demo version (CD-ROM) 
  • Data Transfer Cable
  • EUR Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT EUR)
  • 15 – 20 A Adapter (2719-0154)
  • Null Post Adapter (ESA620-NPA)
  • 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016)
  • Carrying Case (9530-0075) 
  • Power Cord Brazilian (LINE CORD)
TA-ESA615-NA220VESA615 North America, 220 V with test automation Includes:
  • ESA615-10-NA220V
  • Ansur ESA615 Plug-In (ANSUR ESA615)
ESA615-10-NA220VESA615-10 North America 220V Includes:
  • Operator’s Manual (multilingual CD) (CD-ROM) 
  • Getting Started Guide, hard copy, multilingual (Manual)
  • Ansur Plug-in, CD with demo version (CD-ROM)
  • Data Transfer Cable
  • USA Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT USA)
  • Null Post Adapter (ESA620-NPA)
  • 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016)
  • Carrying Case (9530-0075)
  • Power Cord US with C19 Plug (75033) 
  • Power Cord NA 220 V (75057)
  • NA 220 V Adapter Kit (1800001)
TA-ESA615-JAPANESA615 JAPAN, 100 V with test automation Includes: 
  • ESA615-09-JAPAN
  • Ansur ESA615 Plug-In (ANSUR ESA615)
ESA615-09-JAPANESA615-09 Japan, 100 V Includes:
  • Operator’s Manual (multilingual CD) (CD-ROM)
  • Getting Started Guide, hard copy, multilingual (Manual)
  • Ansur Plug-in, CD with demo version (CD-ROM)
  • Data Transfer Cable
  • USA Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT USA)
  • 2195732 15-20A Adapter (2719-0154)
  • Null Post Adapter (ESA620-NPA)
  • 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016)
  • Carrying Case (9530-0075)
  • Power Cord USA (LINE CORD)
TA-ESA615-THAIESA615 THAI, 230V with test automation Includes:
  • ESA615-08-SWISS
  • Ansur ESA615 Plug-In (ANSUR ESA615)
ESA615-08-THAIESA615-08 Thailand, 230 V Includes:
  • Operator’s Manual (multilingual CD) (CD-ROM)
  • Getting Started Guide, hard copy, multilingual (Manual)
  • Ansur Plug-in, CD with demo version (CD-ROM)
  • Data Transfer Cable
  • USA Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT USA)
  • Null Post Adapter (ESA620-NPA)
  • 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016)
  • Carrying Case (9530-0075)
  • Power Cord US with C19 Plug (75033)
TA-ESA615-SWISSESA615 SWISS, 230 V with test automation Includes: 
  • ESA615-07-SWISS
  • Ansur ESA615 Plug-In (ANSUR ESA615)
ESA615-07-SWISSESA615-07 Switzerland, 230 V Includes:
  • Operator’s Manual (multilingual CD) (CD-ROM) 
  • Getting Started Guide, hard copy, multilingual (Manual)
  • Ansur Plug-in, CD with demo version (CD-ROM)
  • Data Transfer Cable
  • EUR Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT EUR)
  • Null Post Adapter (ESA620-NPA)
  • 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016)
  • Carrying Case (9530-0075)
  • Power Cord Switzerland (FBC-ESA620-4420)
TA-ESA615-UKESA615 UK, 230 V with test automation Includes:
  • ESA615-06-UK
  • Ansur ESA615 Plug-In (ANSUR ESA615)
ESA615-06-UKESA615-06 UK, 230 V Includes:
  • Operator’s Manual (multilingual CD) (CD-ROM)
  • Getting Started Guide, hard copy, multilingual (Manual)
  • Ansur Plug-in, CD with demo version (CD-ROM)
  • Data Transfer Cable
  • EUR Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT EUR)
  • Null Post Adapter (ESA620-NPA) 
  • 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016)
  • Carrying Case (9530-0075)
  • Power Cord UK (75024)
TA-ESA615-AUSESA615 Australia, 230 V with test automation Includes:
  • ESA615-05-AUS
  • Ansur ESA615 Plug-In (ANSUR ESA615)
ESA615-05-AUSESA615-05 Australia/China, 230 V Includes:
  • Operator’s Manual (multilingual CD) (CD-ROM)
  • Getting Started Guide, hard copy, multilingual (Manual)
  • Ansur Plug-in, CD with demo version (CD-ROM)
  • Data Transfer Cable
  • USA Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT USA)
  • Null Post Adapter (ESA620-NPA)
  • 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016)
  • Carrying Case (9530-0075)
  • Power Cord Australian (75025)
TA-ESA615-12-INDIA

TA-ESA615-12-India, 250V with test automation includes:

  • ESA615-12-India
  • Ansur ESA615 Plug-In (ANSUR ESA615)
ESA615-12-INDIA

ESA615-12-India, 250V includes:

  • Operator’s Manual (multilingual CD) (CD-ROM)
  • Getting Started Guide, hard copy, multilingual (Manual)
  • Ansur Plug-in, CD with demo version (CD-ROM)
  • Data Transfer Cable
  • EUR Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT EUR)
  • Null Post Adapter (ESA620-NPA)
  • 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016)
  • Carrying Case (9530-0075)
  • Power Cord India
TA-ESA615-ISRESA615 Israel, 230 V with test automation Includes: 
  • ESA615-03-ISR
  • Ansur ESA615 Plug-In (ANSUR ESA615)
ESA615-03-ISRESA615-03 Israel, 230 V Includes:
  • Operator’s Manual (multilingual CD) (CD-ROM)
  • Getting Started Guide, hard copy, multilingual (Manual)
  • Ansur Plug-in, CD with demo version (CD-ROM)
  • Data Transfer Cable
  • USA Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT USA)
  • Null Post Adapter (ESA620-NPA)
  • 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016)
  • Carrying Case (9530-0075)
  • Power Cord Israel (LINE CORD)
TA-ESA615-EURESA615 Europe, 230 V with test automation Includes: 
  • ESA615-02-EUR
  • Ansur ESA615 Plug-In (ANSUR ESA615)
ESA615-02-EURESA615-02 Europe, 230 V Includes:
  • Operator’s Manual (multilingual CD) (CD-ROM)
  • Getting Started Guide, hard copy, multilingual (Manual)
  • Ansur Plug-in, CD with demo version (CD-ROM)
  • Data Transfer Cable
  • EUR Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT EUR)
  • Null Post Adapter (ESA620-NPA)
  • 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016)
  • Carrying Case (9530-0075)
  • Power Cord Europe (75026)
TA-ESA615-FRESA615 France/Belgium, 230 V with test automation Includes:
  • ESA615-01-FR
  • Ansur ESA615 Plug-In (ANSUR ESA615)
ESA615-01-FRESA615-01 France/Belgium, 230V Includes:
  • Operator’s Manual (multilingual CD) (CD-ROM)
  • Getting Started Guide, hard copy, multilingual (Manual)
  • Ansur Plug-in, CD with demo version (CD-ROM)
  • Data Transfer Cable
  • EUR Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT EUR)
  • Null Post Adapter (ESA620-NPA)
  • 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016)
  • Carrying Case (9530-0075)
  • Power Cord Europe (75026)
TA-ESA615-USA

ESA615 US, 115 V with test automation includes:

  • ESA615-USA
  • Ansur ESA615 Plug-In (ANSUR ESA615)
ESA615-USAESA615 US, 115V Includes:

  • Operator’s Manual (multilingual CD) (CD-ROM)
  • Getting Started Guide, hard copy, multilingual (Manual)
  • Ansur Plug-in, CD with demo version (CD-ROM)
  • Data Transfer Cable 
  • USA Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT USA)
  • 15 – 20 A Adapter (2719-0154)
  • Null Post Adapter (ESA620-NPA) 
  • 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016)
  • Carrying Case (9530-0075) 
  • Power Cord US (LINE CORD)

Test Leads, Probes and Clips
3111008USA/AUS/ISR Accessory Kit: test lead set, TP1 test probe set, AC285 alligator clip set (ESA T/L kit, USA, country-specific)
6358Retractable Test Leads (part# 1903307)

Batteries, Chargers, and Power Supply
7505775057 Powercord set NEMA 6-15/C19 250V (part# 2238671)
75033Power cable w/C19 plug, detachable, 120 V/15 A – USA (part# 2238644)

Cables
4034393Data Transfer Cable for use with ProSim 4, ProSim 8, ESA615

Carrying Cases
2248650Carrying Case

Miscellaneous Accessories
2719-0154Power cord adapter 20-15A - (part# 2195732) - Used to adapt a 20 Amp power cord so it will plug into a 15 Amp receptacle.
3111024EUR Accessory Kit: test lead set, TP74 test probe set, AC285 alligator clip set (ESA T/L kit, EUR)
LINE CORDESA615 country-specific Line Cord
1210 ECG1-to-10 ECG Adapter Box Assembly (part# 3392119)
for use with ESA612 and ESA615
ESA612-20165-to-5 Banana Jack to ECG Adapter (BJ2ECG) (part# 3359538)
ESA620-NPA null post assembly (part# 3326842)
Null post adapter to zero test lead resistance
9503-0004Ground Pin Adapter (US receptacle testing ground lug) (part# 2392639)
3472633Ultrasound Test Cable Adapter, optional (part# 3472633)
17024Universal Banana Adapter (10-pack)(part# 2462072)
4151242USA/NEMA outlet to NBR14136 socket (Brazil only)

Software
ANSUR ESA615Ansur ESA615 Plug-in license (part# 4165219)

Service Packages
Care PlansRequest a quote