Impulse 7000DP Defibrillator Tester, External Pacer Analyzer

Ensure proper operation and ultimate performance of critical life-support cardiac-resuscitation equipment, including mono- and bi-phasic defibrillators and AEDs.

Impulse 7000DP Defibrillator Analyzer
Fluke Biomedical defibrillator analyzers ensure proper operation and ultimate performance of critical life-support cardiac-resuscitation equipment, including mono- and bi-phasic defibrillators and AEDs. These best-in-class defibrillators offer test capabilities that encompass a global spectrum of established pulse shapes, feature breakthrough AED technology compatibility, and deliver superior performance with accuracy readings ± 1% of reading +0.1J.

The Impulse 7000DP model is a combination Defibrillator / Transcutaneous Pacemaker Analyzer with an intuitive user interface, backlight, and easy-to-read display. This portable, rugged, easy-to-carry device features functionality that includes energy output measurement, ECG waves, and an extensive range of test loads and measurement algorithms needed to test external transcutaneous pacemakers. The Impulse 7000DP coupled with the Impulse 7010 Defibrillator Selectable Load Accessory enables end-users to select from multiple loads (25Ω - 200Ω) ensuring IEC 60601-2-4 compliance.

The Impulse 7000DP features 10 independent ECG outputs that provide 12 lead combinations for standardized clinical signals.

Ansur test automation may be purchased in order to standardize test procedures, capture waveforms and test results, and evolve with increasing requirements for electronic data storage.

Key Features

Defibrillator test equipment
  • IEC 60601-2-4 Compliance
  • Lown, Edmark, trapezoidal, biphasic and pulsed-biphasic defibrillation technology compatibility
  • AED compatibility
  • 12-lead ECG simulation
  • Ability to test external transcutaneous pacemakers (Impulse 7000 DP only)
  • Internal pacer brand selections
  • Flexible heart-rate settings (1BPM step)
  • DSP-based measurements for firmware and waveform upgrade
  • Waveform capture, store and replay
  • First-in-class accuracy ± 1% of reading + 0.1J
  • Long lasting, rechargeable battery
General Specifications
Temperature, Operating  10 °C to 40 °C (50 °F to 104 °F)
Temperature, Storage  -20 °C to +60 °C (-4 °F to +140 °F)
Humidity  10 % to 90 % non-condensing
Display  LCD display
Communications  USB device port for computer control
Modes of Operation  Manual and remote
Power  Internal rechargeable NiMH battery pack for nine hours (typical) operation after full charge or the battery charger can operate the analyzer and charge the battery simultaneously
Battery Charger  100 V to 240 V input, 15 V/1.5 V A output. For best performance, the battery charger should be connected to a properly grounded ac receptacle
Enclosure  ABS plastic housing
Dimensions  32 cm x 24 cm x 13 cm

 (13 in x 9.5 in x 5 in)
Weight  3.02 kg (6.6 lb, 0.1 oz)
Safety Standards  CE: IEC/EN61010-1 2nd Edition; Pollution degree 2 CAN/CSA-C22.2 No. 61010-1; UL61010-1 C-Tick: Australian EMC

Energy Output Measurement Specifications
Compatible Defibrillator Waveshapes
 Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic
 AC Pulsed Biphasic waveform has not been approved in the United States.
Auto-ranged Measurement
 0.1 J to 600 J
Accuracy
 ± (1 % of reading + 0.1 J)
Load Resistance
Resistance  50 Ω
Accuracy  ±1 %, non-inductive (<2 µH)
Pulse Trigger Level
 20 V
Pulse Width
Range  1 ms to 50 ms
Accuracy  ±0.1 ms
Voltage
Range  20 V to 5000 V
Accuracy  ± (1 % of reading + 2 V)
Current
Range  0.4 A to 100 A
Accuracy  ± (1 % of reading + 0.1 A)
Tilt (Biphasic and Pulsed Biphasic)
Range  1 % to 99 %
Accuracy  ± 1 digit
Interphase Delay (Biphasic and Pulsed Biphasic)
Range  0.1 ms to 9.9 ms
Accuracy  ± 0.1 ms
Frequency (Pulsed Biphasic Only)
Range  2000 Hz to 8000 Hz
Accuracy  ± 1 % of reading
Duty Cycle (Pulsed Biphasic Only)
Range  1 % to 99 %
Accuracy  ± digit
Sample Rate
 250 kHz (4 µs sample)
Maximum Average Power
 12 W, equivalent to 10 defib pulses of 360 J every 5 minutes
Scope Output
 Autorange:2000:1, 400:1 and 80:1: dependant on the range
Waveform Playback
 
  • Output: BNC
  • Output impedance: 50 Ω
  • Amplitude Accuracy: ± 5 %
Charge Time Measurement
Range  0.1 s to 100 s
Accuracy  ± 0.05 s, typical
Synchronization Test (Elective Cardioversion)
Delay Time Measurement  Timing window: : ECG R-wave peak to the defib pulse peak
 Range: -120 to +380 ms; measures timing from 120 ms prior to the R-wave peak to up to 380 ms following the R-wave peak
 Resolution: 1 ms
 Accuracy: ± 1 ms
ECG waves  Normal Sinus Rhythm (NSR): 10 BPM to 180 BPM, in 1 BPM steps
 Atrial fibrillation: Coarse and fine
 Monomorphic Ventricular Tachycardia: 120 BPM to 240 BPM, in 5 BPM steps
 Asystole: Flat line
Automated Defibrillator Test ECG Waves
Normal Sinus  10 BPM to 300 BPM, in 1 BPM steps
Ventricular Fibrillation  Coarse and fine
Monomorphic Ventricular Tachycardia  120 BPM to 300 BPM, in 1 BPM steps
Polymorphic Ventricular Tachycardia  5 types
Asystole  Flat line

 

ECG Waves
ECG General
Lead configuration  12-lead simulation. RA, LL, LA, RL, V1-6 with independent signal outputs
Lead-to-lead impedance  1000 Ω (nominal)
Rate accuracy  ± 1 % of nominal
ECG Amplitudes
Reference lead  Selectable, Lead II (default) or Lead 1
Settings  0.05 mV to 0.45 mV, in 0.05 mV steps
0.5 mV to 5 mV, in 0.5 mV steps
Accuracy (All Performance waves and Normal Sinus R waves)  Lead II: ± 2 %
All other leads: ± 5 %
Defib paddles: ± 5 %
Amplitude of ECG signals relative to amplitude setting (in percent)  Lead II reference:
 Performance waves and R wave detection
 Lead I – 70 %
 Lead II – 100 %
 Lead III – 30 %
 Lead V1 – 100%
 Lead V2 – 100%
 Lead V3 – 100 %
 Lead V4 – 100 %
 Lead V5 – 100 %
 Lead V6 – 100 %
 Normal Sinus waves
 Lead I – 70 %
 Lead II – 100 %
 Lead III – 30 %
 Lead V1 – 24%
 Lead V2 – 48%
 Lead V3 – 100 %
 Lead V4 – 120 %
 Lead V5 – 112 %
 Lead V6 – 80 %
 Lead I reference:
 Performance waves and R wave detection
 Lead I – 100 %
 Lead II – 150 %
 Lead III – 50 %
 Lead V1 – 100%
 Lead V2 – 100%
 Lead V3 – 100 %
 Lead V4 – 100 %
 Lead V5 – 100 %
 Lead V6 – 100 %
 Normal Sinus waves
 Lead I – 100 %
 Lead II – 150 %
 Lead III – 50 %
 Lead V1 – 24%
 Lead V2 – 48%
 Lead V3 – 100 %
 Lead V4 – 120 %
 Lead V5 – 112 %
 Lead V6 – 80 %
ECG Normal Sinus
Rates  10 BPM to 360 BPM, in 1 BPM steps
ECG High Level Output (BNC Jack)
Amplitude Range  0.5 V/mV of reference lead setting
Accuracy  ± 5 %
Output Impedance:  50 Ω
ECG On Cefibrillator Input Load
 Same as the Lead II amplitude but limited to ± 4 mV
ECG Performance Waves
Square wave  2 Hz and 0.125 Hz
Triangular wave  2 Hz and 2.5 Hz
Sine waves  0.05, 0.5, 5, 10, 40, 50, 60, 100, 150, and 200 Hz
Pulse  30BPM and 60 BPM, 60 ms pulse width
R-Wave Detection
Waveform  Haver-triangle
Rates  30, 60, 80, 120, 200 and 250 BPM
Widths  8, 10, 12 and 20 to 200 (by 10) ms
Amplitudes  0.05 mV to 0.45 mV 0.05 mV hard return
 0.5 mV to 5 mV, in 0.5 mV steps
Accuracy  ± (1 % setting + 1 ms)
Noise Immunity
Wave  Sine
Line Frequency  50 Hz or 60 Hz ( ± 0.5 Hz)
Amplitude  Range: 0.0 mV to 10 mV, in 0.5 mV steps
 Accuracy: ± 5 %
Transvenous Pacer Pulse Simulation
Widths  Range: 0.1, 0.2, 0.5, 1, and 2 ms.
 Accuracy: ± 5 % of setting
Amplitudes  Range: 0 (off) and ± 2, ± 4, ±6, ±8, ±10, ±12. ±14, ±16, ±18, ±20, ±50, ±100, ±200, ±500, ±700 mV
 Accuracy: ± (10 % setting + 0.2 mV)
Amplitude of Transvenous Pacer Pulse Simulation signals relative to amplitude setting (in percent)  Lead II reference:
 Lead I – 67 %
 Lead II – 100 %
 Lead III – 33 %
 Lead V1 – 67%
 Lead V2 – 67%
 Lead V3 – 67 %
 Lead V4 – 67 %
 Lead V5 – 67 %
 Lead V6 – 67 %
 Lead I reference:
 Lead I – 100 %
 Lead II – 150 %
 Lead III – 50 %
 Lead V1 – 100%
 Lead V2 – 100%
 Lead V3 – 100 %
 Lead V4 – 100 %
 Lead V5 – 100 %
 Lead V6 – 100 %
Arrhythmia Selections
Pacer Interactive (Transcutaneous Pacer, Impulse 7000DP only)  Demand: 30 BPM to 360 BPM, in 1 BPM steps
Asynchronous
Non-Capture
Non-Function
Threshold (interactive pacing simulation only): 10 mA to 250 mA in 10 mA steps
Supraventricular  Atrial Fibrillation Coarse
Atrial Fibrillation Fine
Atrial Flutter
Sinus Arrhythmia
Missed Beat
Atrial Tachycardia
Paroxysmal Atrial Tachycardia
Nodal Rhythm
Supraventricular Tachycardia
Premature  Atrial PAC
Nodal PNC
PVC1 Left Ventricle (LV)
PVC1 Left Ventricle (LV) Early
PVC1 Left Ventricle (LV) R on T
PVC2 Right Ventricle (RV)
PVC2 Right Ventricle (RV) Early
PVC2 Right Ventricle (RV) R on T
Multifocal PVCs (LV and RV)
Ventricular  PVCs 6/min
PVCs 12/min
PVCs 24/min
Freq Multifocal
Trigeminy
Bigeminy
Pair PVCs
Run 5 PVCs
Run 11 PVCs
Monomorphic VTach: 120 to 300 (5 BPM steps)
Polymorphic VTach: 5 selections
Ventricular Fibrillation: Coarse and Fine
Asystole
Conduction  1° Block
2° Block Type I
2° Block Type II
3° Block
Right Bundle Branch Block (RBBB)
Left Bundle Branch Block (LBBB)
Transvenous Pacemaker (ECG simulation only)  Atrial 80 BPM
Asynchronous 75 BPM
Demand with frequent sinus beats
Demand with occasional sinus beats
AV Sequential
Non-Capture
Non-Function
Selectable Pacer pulse Parameters for Transvenous Simulation (Atrial & Ventricular channels are independently selectable by the user)
Atrial Pacer Pulse (Note: relates only to the TV pacer)  Width: 0.1, 0.2, 0.5, 1, 2 ms
Polarity: + or -
Amplitude: 0 (off), 2 to 20 (by 2), 50, 100, 200, 500, 700 mV
Ventricular Pacer Pulse (Note: relates only to the TV pacer)  Width: 0.1, 0.2, 0.5, 1, 2 ms
Polarity: + or -
Amplitude: 0 (off), 2 to 20 (by 2), 50, 100, 200, 500, 700 mV
(BNC Jack)
Accuracy  ± 5 % for 2 Hz Square Wave , 50Ω output impedance

7000DP Specifications
Test Load Selections
Defibrillator Input  Fixed load: 50 Ω
Accuracy: ± 1 %, non-inductive (<2 µH)
Power rating: 10 defib pulses of 360 J every 5 minutes
Pacemaker Input  Variable load: 50 Ω to 1500 Ω in 50 Ω steps
Accuracy: ± 2 %, non-inductive (< 2 µH)
Power rating: 5 W (average), 40 W (peak) @ 1000 Ω
Measurements
Manufacturer specific algorithms  GE Responder (1500 & 1700)
MDE 300 (Medical Data Electronics)
Medtronic ERS/Physio Control LIFEPAK
MRL (Medical Research Laboratory/Welch Allyn)
Philips/Agilent HP
Schiller Medical
ZOLL Medical
(plus a general purpose Default Algorithm selection)
Current  4 to 250 mA, ± (1% of reading + 0.02 mA)
Pulse rate  5 to 800 PPM, ± (0.5% of reading + 0.1 PPM)
Pulse width  1 to 100 ms, ± (0.5% of reading + 0.01 ms)
Energy  1 ?J to 2 J, ± (4% of reading + 10 ?J)
Demand and Asynchronous Mode Test
Input pacer pulse rates  30 to 200 PPM
ECG NSR wave  Rate:10 to 300 BPM in 1 BPM increments.
Amplitude:1 mV
Underdrive rate:10 BPM minimum
Overdrive rate:300 BPM maximum
Sensitivity Test
Automatic interactive threshold detection  Compatible pacer rates: 30 to 120 PPM.
ECG R Wave  Waveforms:Square, Triangle, Sine
Width:1 to 19 (by 1) ms. 20 to 95 (by 5) ms. 100 to 300 (by 25) ms
Amplitude:0.05 to 0.95 (by 0.05) mV. 1 to 5 (by 0.5) mV
Accuracy:± 5% of setting
Refractory Period Tests
Paced Refractory Period  20 to 500 ms
Sensed Refractory Period  15 to 500 ms
Accuracy  ± 1 ms
Pacer pulse rate  20 to 200 PPM
ECG  Waveform:Triangle wave
Pulse Width:40 ms
Amplitude:1 mV

Model NameDescription
Impulse 7000DP Impulse 7000DP Defibrillator/Transcutaneous Pacemaker Analyzer 120 V - US

All models include:
  • USB Computer Communication Cable (Cable assembly)
  • User Manual CD (CD-ROM)
  • Getting Started Guide (Manual)
  • Battery Eliminator (country specific)
  • Defib Paddle Contact Plates (FBC-IMP7000-8010)
  • Carrying Case (Soft case)
Impulse 7000DP-01Impulse 7000DP Schuko
Impulse 7000DP-02Impulse 7000DP UK
Impulse 7000DP-03Impulse 7000DP Japan
Impulse 7000DP-04Impulse 7000DP Australia
Impulse 7000DP-05Impulse 7000DP India
Impulse 7000DP Defibrillator/Transcutaneous Pacemaker Analyzer BrazilImpulse 7000DP Brazil
TA-IMP7KDPImpulse 7000DP with test automation 120 V US

All models include:
  • USB Computer Communication Cable (Cable assembly)
  • User Manual CD (CD-ROM)
  • Getting Started Guide (Manual)
  • Battery Eliminator (country specific)
  • Defib Paddle Contact Plates (FBC-IMP7000-8010)
  • Carrying Case (Soft case)
TA-IMP7KDP-01Impulse 7000DP with test automation Shuko
TA-IMP7KDP-02Impulse 7000DP with test automation United Kingdom
TA-IMP7KDP-03Impulse 7000DP with test automation Japan
TA-IMP7KDP-04Impulse 7000DP with test automation Australia
TA-IMP7KDP-05Impulse 7000DP with test automation India
Impulse 7000DP Defibrillator/Transcutaneous Pacemaker Analyzer with test automation BrazilImpulse 7000DP with test automation Brazil
TA-IMP7K/7010 USImpulse 7000DP US 120 V with Impulse 7010 and test automation United States

All models include
  • USB Computer Communication Cable (CABLE ASSEMBLY)
  • User Manual CD (CD-ROM)
  • Getting Started Guide (Manual)
  • Battery Eliminator (country specific)
  • Defib Paddle Contact Plates (FBC-IMP7000-8010)
  • Carrying Case (Soft Case)
  • Ansur Test-Automation Software (ANSUR IMPULSE 7000)
TA-IMP7K/7010 SHKImpulse 7000DP with Impulse 7010 and test automation Shuko
TA-IMP7K/7010 UKImpulse 7000DP with Impulse 7010 and test automation United Kingdom
TA-IMP7K/7010 JPNImpulse 7000DP with Impulse 7010 and test automation Japan
TA-IMP7K/7010 AUSImpulse 7000DP with Impulse 7010 and test automation Australia
Impulse 7000DP with Impulse 7010 and test automation BrazilImpulse 7000DP with Impulse 7010 and test automation Brazil

Defibrillator Adapters
I6/7 D/P ADPT102Internal discharge paddle contacts (2) (4mm) (part# 3065438)
I6/7 D/P ADPT106MedtronicERS/Physio-Control (part# 3065489)
(FAST PATCH) (set of two) 4mm
for Impulse 6000D/7000DP

Pacer Adapters
I6/7 D/P ADPT109Zoll Medical PD2200 Multi-Function Series electrodes (part# 3065511)
M Series, M Series-CCT, R Series, E Series, AED Pro™ and AED Plus™ (Includes inline connector) (defib and pace) 4mm (AED Pro and AED Plus requires an auxiliary adapter part number: 8000-0804-01 available from Zoll Medical)

Defibrillator Analyzers
Impulse 7010Impulse 7010 Defibrillator Selectable Load Accessory (*compatible with Impulse 7000 only)

Adapters
I6/7 D/P ADPT101GE Marquette defib/pacer adapter (part# 3065423)
(RESPONDER1500/1700 Series) (set of two) 4mm
I6/7 D/P ADPT103Kimberly Clark/R2 Darox MRL/MDE/NK: 4 mm defibrillator adapters (part# 3065450)
I6/7 D/P ADPT104Medtronic ERS/Physio-Control Defib/Pacer adapter (part# 3065461)
(QUIK COMBO) 4mm
I6/7 D/P ADPT105Medtronic ERS/Physio-Control pacer adapter (part# 3065477)
(QUIK PACE) (set of two): 4 mm
I6/7 D/P ADPT108Philips/Agilent Defib/Pacer adapter (part# 3065509)
HEARTSTART FR2/MRX 4mm
I6/7 D/P ADPT107Philips/Agilent/HP Defib/Pacer adapter (part# 3065492)
(CODEMASTER Series-Round) 4mm
I6/7 D/P ADPT110Zoll Medical NTP/PD1400:4 mm pacer adapters (part# 3065527)

Software
Ansur IMPULSE 7000Ansur Impulse 6000D/7000DP Plug-In software (part# 3091370)

Service Packages
Care PlansRequest a quote
Biomedical Equipment  |  Patient Simulator  |  Safety Analyzer  |  Dosimeter  |  Radiation Detector
Search Fluke Biomedical
ResourcesProducts
Find Resources
Related Products
View related products
 
 Impulse 6000D Defibrillator Analyzer »
 
 Impulse 7010 Defibrillator Selectable Load Accessory »
 
 SigmaPace™ 1000 External Pacemaker Analyzer  »

Impulse 6000D/7000DP Defibrillator/External Pacer Analyzer and Tester Data Sheet


Data Sheet