ProSim 8 Vital Signs Patient Simulator

PMs done in five minutes
ProSim 8 Patient Simulator See video »

Learn about the ProSim 4 »

ProSim 8 Vital Signs Patient Simulator
ProSim 8 Vital Signs Cardiac Simulator
ProSim 8 Vital Signs Heart Simulator
ProSim 8 Vital Signs ECG Simulator
ProSim 8 Vital Signs EKG Simulator

The ProSim 8; a pulse oximeter, patient simulator, ecg simulator, and more

Designed to make comprehensive patient monitor testing fast and easy, ProSim 8 Vital Signs Simulator offers total patient monitor quality and safety testing from non-invasive blood pressure (NIBP) static pressure linearity and dynamic pressure repeatability tests, EC13 and ACLS ECG waveforms, to complete Nellcor and Nonin SpO2 low perfusion and Masimo Rainbow SET. Featuring specialized stay-connected ECG posts to ensure secure lead connections, ProSim 8 is the preferred tester for patient monitor quality assurance and safety professionals.

The ProSim 8 tests ECG (including fetal ECG/IUP and arrhythmias), respiration, temperature, IBP/cardiac catheterization, cardiac output, NIBP and SpO2 (including Masimo multi-wavelength Rainbow SET)s in a single 5-minute PM testing tool. Wireless PC communication, customized presets, auto-sequences, barcode scanning, onboard memory, direct data capture, printing functions and single-step adjustability maximizes testing productivity; a customized carry case doubles as a mobile work station.

Multi-language user interface available in English, Spanish, German, French, Japanese, Italian, and Chinese.

  • All-in-one patient simulator 80 % smaller and 17 lbs./7.7 kilos lighter than single function simulator test tools
  • 8-in-1 multifunction patient simulator is a combination ECG simulator, fetal simulator, arrhythmia simulator, respiration simulator, temperature simulator, IBP simulator, cardiac output simulator, cardiac catheterization simulator, NIBP simulator, SpO2 tester, and is the premier SpO2 simulator to test Rainbow multi-wavelength waveforms
  • Stay-connected ECG posts for easy/secure ECG snap and lead connections
  • Custom SpO2 r-curve for accurate testing of the latest and future oximetry technologies
  • Static pressure linearity testing
  • Repeatable NIBP simulation for dynamic pressure repeatability testing
  • Physiologically synchronized pulses across all parameters
  • Barcode scanning and direct data capture and printing functionality
  • Onboard, customizable patient pre-sets and auto sequences for fast/easy testing
  • Multi-language user interface offers choice of language selection
  • Integrated, easily-replaceable long-life battery
  • Optional PC-interface software offers customizable procedures/checklists and automated data capture/storage
  • Wireless communication for remote PC control of test device, as well as data transfer and automated regulatory reporting
  • General Specifications
    Temperature
    Operating: 10 °C to 40 °C (50 °F to 104 °F)
    Storage: -20 °C to +60 °C (- 4 °F to 140 °F)
    Humidity 10 % to 90 % non-condensing
    Altitude 3,000 meters (9,843 ft)
    Dimensions (L x W x H) 14.5 cm x 30.2 cm x 8.6 cm (5.7 in x 11.9 in x 3.4 in)
    Display LCD color display
    Communication
    USB device upstream port Mini-B connector for control by a computer
    USB host controller port Type A, 5 V output, 0.5 A max load. Connector for keyboard, barcode reader, and printer
    Wireless IEEE 82.15.4 for control by a computer
    Power Lithium-ion rechargeable battery
    Battery Charger: 100 V to 240 V input, 15 V/2.0 A output. For best performance, the battery charger should be connected to a properly-grounded ac receptacle
    Battery Life 9 hours (minimum), 100 NIBP cycles typical
    Weight 1.87 kg (4.2 lb)
    Safety Standards IEC/EN61010-1 3rd Edition; Pollution degree 2 CAT None
    Certifications CE, CSA, C-TICK N10140 , RoHS
    Electromagnetic Compatibility (EMC) IEC 61326-1:2006

     

    Detailed Specifications
    Normal-Sinus-Rhythm Waveform
    ECG Reference The ECG amplitudes specified are for Lead II (calibration), from the baseline to the peak of the R wave. All other leads are proportional.
    Normal Sinus Rhythm 12-lead configuration with independent outputs referenced to right leg (RL) Output to 10 Universal ECG Jacks, color-coded to AHA and IEC standards.
    High-Level Output 0.2 V/mV ± 5 % of the ECG amplitude setting available on a BNC connector.
    Amplitude 0.05 mV to 0.5 mV (0.05 mV steps) 0.5 mV to 5 mV (0.25 mV steps)
    Other leads are proportional to Lead II (reference lead) in percentage per:
    Lead I: 70
    Lead II: 100
    Lead III: 30
    Lead V1: 24
    Lead V2: 48
    Lead V3: 100
    Lead V4: 120
    Lead V5: 112
    Lead V6: 80
    Amplitude Accuracy ± (2 % of setting + 0.05 mV)
    ECG Rate 10 BPM to 360 BPM in 1 BPM steps
    Rate Accuracy ±1 % of setting
    ECG Waveform Selection Adult (80 ms) or neonatal (40 ms) QRS duration
    ST-Segment Elevation Adult mode only. -0.8 mV to +0.8 mV (0.1 mV steps). Additional steps: + 0.05 mV and - 0.05 mV
    Power-On Default 60 BPM, 1.0 mV, adult QRS and ST-segment elevation of 0 mV
    Arrhythmia
    Pacer Pulse
    Amplitude 0 (off), ± 2, ± 4, ± 6, ± 8, ± 10, ± 12, ± 14, ± 16, ± 18, ± 20, ± 50, ± 100, ± 200, ± 500, and ± 700 mV for lead II (reference lead)
    Accuracy Reference lead II: ± (5 % setting + 0.2 mV)
    All other leads: ± (10 % setting + 0.4 mV)
    Pacer Pulse Width 0.1 ms, 0.2 ms, 0.5 ms, 1 ms, and 2 ms ± 5 %
    Paced Arrhythmias Atrial 80 BPM
    Asynchronous 75 BPM
    Demand with frequent sinus beats
    Demand with occasional sinus beats
    Atrio-ventricular sequential
    Noncapture (one time)
    Nonfunction
    Power-On Default Amplitude 5 mV, width 1 ms, atrial waveform
    Arrhythmia
    Baseline NSR 80 BPM
    PVC Focus Left focus, standard timing (except where specified)
    Supraventricular Arrhythmia Atrial fibrillation (coarse or fine); atrial flutter; sinus arrhythmia; missed beat (one time); atrial tachycardia; paroxysmal atrial tachcardia; nodal rhythm; and supraventricular tachycardia
    Premature Arrhythmia Premature atrial contraction (PAC); premature nodal contraction (PNC); PVC1 left ventricular; PVC1 left ventricular, early; PVC1 left ventricular, R on T; PVC2 right ventricular; PVC2 right ventricular, early; PVC2 right ventricular, R on T; and multifocal PVCs
    Ventricular Arrhythmia PVCs 6, 12, or 24 per minute; frequent multifocal PVCs; bigeminy; trigeminy; multiple PVCs (one-time run of 2, 5, or 11 PVCs); mono-ventricular tachycardia (120 to 300 BPM in 5 BPM steps); poly-ventricular tachycardia (5 types); ventricular fibrillation (coarse or fine); and asystole
    Conduction Defect First-, second-, or third-degree heart block; and right- or left-bundle-branch block
    Advanced Cardiac Life Support
    Shockable pulseless arrest rhythms Ventricular fibrillation (coarse), ventricular fibrillation (fine), unstable polymorphic ventricular tachycardia
    Non-shockable pulseless arrest rhythms Asystole
    Symptomatic bradycardia Sinus bradycardia (< 60 BPM)
    2nd degree AV block, mobitz type I
    2nd degree AV block, mobitz type II
    Complete/3rd degree AV block
    Right bundle branch block
    Left bundle branch block
    Symptomatic tachycardia: regular narrow-complex tachycardia (QRS < 0.12 seconds) Sinus tachycardia > 150 BPM
    Supraventricular Tachycardia
    Symptomatic tachycardia: regular wide-complex tachycardias (QRS > 0.12 seconds) Sinus tachycardia > 150 BPM
    Supraventricular Tachycardia SVT with aberrancy
    Irregular tachycardia Atrial fibrillation (coarse and fine), atrial flutter, unstable monomorphic ventricular tachycardia (120 BPM to 300 BPM), torsade de pointes/polymorphic ventricular tachycardia (long QT interval)
    ECG Performance Testing
    Amplitude (peak-to-peak) 0.05 mV to 0.5 mV (0.05 mV steps) 0.5 mV to 5 mV (0.25 mV steps)
    Other leads are proportional to Lead II (reference lead) in percentage per:
    Lead I: 70
    Lead II: 100
    Lead III: 30
    Lead V1 through V6: 100
    Pulse Wave 30 BPM, 60 BPM, with 60 ms pulse width
    Square Wave 0.125 Hz, 2 Hz, 2.5 Hz
    Triangle Wave 0.125 Hz, 2 Hz, 2.5 Hz
    Sine Wave 0.05 Hz, 0.5 Hz, 1, 2 Hz, 5 Hz, 10 Hz, 25 Hz, 30 Hz, 40 Hz, 50 Hz, 60 Hz, 100 Hz, and 150 Hz
    R-Wave Detection
    Waveform Triangular pulse
    Rate 30 BPM, 60 BPM, 80 BPM, 120 BPM, 200 BPM, and 250 BPM
    Width 8 ms to 20 ms in 2 ms steps, and 20 ms to 200 ms in 10 ms steps
    Width accuracy ± (1 % of setting + 1 ms)
    QRS Detection
    Widths 8 ms to 20 ms in 2 ms steps, and 20 ms to 200 ms in 10 ms steps
    Width accuracy ± (1 % of setting + 1 ms)
    Rate 30 BPM, 60 BPM, 80 BPM, 120 BPM, 200 BPM, and 250 BPM
    R-Wave up slope 0.875 amplitude, 0.4375 x width
    R-Wave down slope Full amplitude, 0.5 x width
    S-Wave up slope 0.125 amplitude, 0.0625 x width
    Tall T-Wave Rejection
    Waveform QT Interval 350 ms
    T-Wave width 180 ms
    T-Wave shape ½ sinewave
    Amplitude 0 % to 150 % reference lead amplitude in 10 % steps
    Rate 80 BPM
    Rate Accuracy ± 1 % of setting
    Amplitute Accuracy ± (2 % of setting + 0.05 mV)
    ECG Artifact
    Type 50 Hz, 60 Hz, muscular, baseline wander, respiration
    Size 25 %, 50 %, 100 % of the normal sinus R-Wave for each lead
    Lead Select All, RA, LL, LA, V1, V2, V3, V4, V5, V6
    Fetal/Maternal ECG
    Fetal Heart Rate (fixed) 60 BPM to 240 BPM in 1 BPM steps
    Fetal Heart Rate (IUP) 140 BPM at beginning, then varies with pressure
    Intrauterine-Pressure Waveforms Variable deceleration, early deceleration, late deceleration, and uniform acceleration
    Wave Duration 90 seconds, bell-shaped pressure curve, from 0 mmHg to 90 mmHg and returning to 0
    IUP Period 2 min, 3 min, or 5 minutes; and manual
    Default Settings FHR 120 BPM, uniform deceleration wave, manual
    Invasive Blood Pressure
    Channels 2, each independently settable with identical parameters and are individually electrically isolated from all other signals
    Input/Output Impedance 300 Ω - or ± 10 %
    Exciter Input Range 2 to 16 V peak
    Exciter-Input Frequency Range DC to 5000 Hz
    Transducer Sensitivity 5 (default) or 40 µV/V/mmHg
    Pressure Accuracy ± (1 % of setting + 1 mmHg) accuracy guaranteed for dc excitation only
    Static Pressure - 10 to + 300 mmHg in 1 mmHg steps
    Pressure Units mmHg or Kpa
    Dynamic Waveforms
    Types (default pressures Arterial (120/80)
    Radial artery (120/80)
    Left ventricle (120/00)
    Right ventricle (25/00)
    Pulmonary artery (25/10)
    Pulmonary-artery wedge (10/2)
    Right atrium (central venous or CVP) (15/10)
    Pressure variability Systolic and diastolic pressures are independently variable in 1 mmHg steps
    Swan-Ganz Sequence Right atrium, right ventrical (RV), pulmonary artery (PA), pulmonary artery wedge (PAW)
    Cardiac Catheterization
    Chambers Aortic, pulmonary valve, and mitral valve
    Respiration Artifact
    Arterial, radial artery, and left ventricle 5 % to 10 % multiplication
    Other 5 mmHg or 10 mmHg
    BP Output Circular DIN 5-Pin
    Power-On Default 0 mmHg
    Respiration
    Rate 0 (OFF), 10 BrPM to 150 BrPM in 1 BrPM steps
    Waves Normal or ventilated
    Ratio (inspiration:expiration)
    Normal 1:1, 1:2, 1:3, 1:4, 1:5
    Ventilated 1:1
    Impedance Variations (? Ω) 0.00 Ω to 1.00 Ω iin 0.05 Ω steps and 1Ω to 5 Ω in 0.25 Ω steps
    Accuracy Delta ± (3 % of setting + 0.05 Ω)
    Baseline 500 Ω, 1000 Ω (default), 1500 Ω, 2000 Ω, Leads I, II, III
    Accuracy Baseline ±5 %
    Respiration Lead LA or LL (default)
    Apnea Selection 12 sec, 22 sec, or 32 seconds (one-time events), or continuous (Apnea ON = respiration OFF)
    Power-On Default 20 BrPM, delta 1.0 Ω
    Temperature
    Temperature 30 °C to 42.0 °C in 0.5 °C steps
    Accuracy ± 0.4 °C
    Compatibility Yellow Springs, Inc. (YSI) Series 400 and 700
    Output Circular DIN 4-Pin
    Cardiac Output
    Catheter Type Baxter Edwards, 93a-131-7f
    Calibration Coeffecient 0.542 (0 °C injectate), 0.595 (24 °C injectate)
    Blood Temperature 36 °C (98.6 °F) to 38 °C (100.4 °F) ± 2 % in 1 °C steps
    Injectate Volume 10 cc
    Injectate Temperature 0 °C or 24 °C
    Cardiac Output 2.5, 5, 10 liters per minute ± 7.5 %
    Faulty-Injectate Curve Waveform for simulation available
    Left-To-Right-Shunt Curve Waveform for simulation available
    Calibrated Pulse 1.5 ° for 1 second
    Connector Circular DIN 7 pin
    Power-On Default 5 liters per minute, 0 °C injectate, 37 °C blood temperature
    Non-Invasive Blood Pressure
    Pressure Units mmHg or kPa
    Manometer (Pressure Meter)
    Range 10 mmHg to 400 mmHg
    Resolution 0.1 mmHg
    Accuracy ± (0.5 % reading + 0.5 mmHg)
    Pressure Source
    Target pressure range 20 mmHg to 400 mmHg
    Resolution 1 mmHg
    NIBP Simulations
    Pulse 2 mmHg max into 500 ml NIBP system
    Volume of air moved 1.25 ml max
    Simulations (systolic/diastolic [MAP]) Adult: 60/30 (40), 80/50 (60); 100/65 (77); 120/80 (93); 150/100 (117); and 200/150 (167) and 255/195 (215)
    Neonatal: 35/15 (22); 60/30 (40); 80/50 (60); 100/65 (77); 120/80 (93) and 150/100
    Pressure variability: systolic and diastolic pressures are variable by 1 mmHg
    Repeatability Within ± 2 mmHg (at maximal pulse size independent of device under test)
    Synchronization: normal Sinus heart rates: 30 BPM to 240 BPM Maximum rate at 1 ml: 240 BPM achievable with pulses up to 1 ml
    Maximum rate at 1.25 ml: 180 BPM
    Synchronization: Arrhythmias Premature atrial contraction (PAC), premature ventricular contraction (PVC), atrial fibrillation, and missed beat
    Leak Test
    Target pressure 20 to 400 mmHg
    Elapse time 0:30 to 5:00 minutes: seconds in 30 second steps
    Leakage rate 0 mmHg/minute to 200 mmHg/minute
    Pressure Relief Test Range 100 to 400 mmHg
    SpO2 Test (Optional)
    % O2
    Range 30 % to 100 %
    Resolution 1 %
    % O2 Accuracy
    With oximeter manufacturer's R-curve Saturation within UUT specific range: ± (1 count + specified accuracy of the UUT)
    Saturation outside UUT specific range: monotonic with unspecified accuracy
    With Fluke Biomedical R-curves 91 % to 100 % ± (3 counts + specified accuracy of the UUT)
    81 % to 90 % ± (5 counts + specified accuracy of the UUT)
    71 % to 80 % ± (7 counts + specified accuracy of the UUT)
    Below 7 % monotonic with unspecified accuracy
    Heart Rate 30 BPM to 300 BPM in 1 BPM steps. SpO2 test is synchronized with ECG rate delayed by 150 ms, accuracy ±1% of setting
    Transmission: ratio of detector current to LED current, expressed in parts per million (ppm)
    Range 0 ppm to 300.00 ppm
    Resolution 0.01 ppm
    Accuracy + 50 %/- 30 % for compatible monitors, unspecified for others. Selected by finger size and color: dark, thick finger, medium finger, light, thin finger, neonatal foot. The full range and resolution are available in the engineering mode
    Pulse Amplitude
    Range 0 % to 20.00 %
    Resolution 0.01 %
    Artifact
    Respiration Range: 0 % to 5 % of transmission
    Resolution: 1 %
    Rate: all ProSim respiration simulation settings
    Ambient light Range: 0 to 5X transmitted light
    Resolution: 1X
    Frequency: DC, 50 Hz, 60 Hz, and 1 kHz to 10 kHz in 1 kHz steps
    Masimo Rainbow Technology Test Masimo Rainbow technology with an optional adapter supplied by Masimo that allows the ProSim two wavelength to test the Rainbow multiple wavelength system
    Compatible Manufacturer Products
    With manufacturer R-curve Nellcor, Masimo, Nonin, and Nihon Kohden
    With Fluke R-curve Mindray, GE-Ohmeda, Philips/HP, and BCI
    Pre-Defined Simulations
    Normal
    Hypertensive
    Hypotensive
    Tachycardic
    Bradycardic
    Ventricular Fibrillation
    Asystole
    Autosequences
    Monitor Testing Sequence
    ECG Sequence
    Oximeter Testing Sequence
    Cardiac Failure Sequence
    Arrhythmia Sequence
    Exercise Sequence
    Respiration Sequence
    NIBP Testing Sequence
    IBP Testing Sequence
    Temperature Sequence
    Model NameDescription
    ProSim 8
    (3979409)
    ProSim 8 Vital Signs Simulator US Includes:
    • ProSim 8 Getting Started Manual (3980667)
    • ProSim 8 User's Manual (3980671)
    • ProSim 8 Battery Pack (4021085)
    • USB Cable (4034393)
    • IBP Cable, unterminated (2392173)
    • ProSim 8 Carrying Case (4034597)
    • ProSim NIBP MANDREL SET (4308086)
    • AC/DC Power Supply
    • AC Power cord USA
    • Ansur demo CD-ROM
    ProSim 8-01
    (3979411)
    ProSim 8 Vital Signs Simulator Schuko Includes:
    • ProSim 8 Getting Started Manual (3980667)
    • ProSim 8 User's Manual (3980671)
    • ProSim 8 Battery Pack (4021085)
    • USB Cable (4034393)
    • IBP Cable, unterminated (2392173)
    • ProSim 8 Carrying Case (4034597)
    • ProSim NIBP MANDREL SET (4308086)
    • AC/DC Power Supply
    • AC Power cord Schuko
    • Ansure Demo CD-ROM
    ProSim 8-02
    (3979427)
    ProSim 8 Vital Signs Simulator UK Includes:
    • ProSim 8 Getting Started Manual (3980667)
    • ProSim 8 User's Manual (3980671)
    • ProSim 8 Battery Pack (4021085)
    • USB Cable (4034393)
    • IBP Cable, unterminated (2392173)
    • ProSim 8 Carrying Case (4034597)
    • ProSim NIBP MANDREL SET (4308086)
    • AC/DC Power Supply
    • AC Power cord UK
    • Ansur demo CD-ROM
    ProSim 8-03
    (3979430)
    ProSim 8 Vital Signs Simulator Japan Includes:
    • ProSim 8 Getting Started Manual (3980667)
    • ProSim 8 User's Manual (3980671)
    • ProSim 8 Battery Pack (4021085)
    • USB Cable (4034393)
    • IBP Cable, unterminated (2392173)
    • ProSim 8 Carrying Case (4034597)
    • ProSim NIBP MANDREL SET (4308086)
    • AC/DC Power Supply
    • AC Power cord Japan
    • Ansur demo CD-ROM
    ProSim 8-04
    (3979448)
    ProSim 8 Vital Signs Simulator Australia Includes:
    • ProSim 8 Getting Started Manual (3980667)
    • ProSim 8 User's Manual (3980671)
    • ProSim 8 Battery Pack (4021085)
    • USB Cable (4034393)
    • IBP Cable, unterminated (2392173)
    • ProSim 8 Carrying Case (4034597)
    • ProSim NIBP MANDREL SET (4308086)
    • AC/DC Power Supply
    • AC Power cord Australia
    • Ansur demo CD-ROM
    ProSim 8-05
    (3979453)
    ProSim 8 Vital Signs Simulator Brazil Includes:
    • ProSim 8 Getting Started Manual (3980667)
    • ProSim 8 User's Manual (3980671)
    • ProSim 8 Battery Pack (4021085)
    • USB Cable (4034393)
    • IBP Cable, unterminated (2392173)
    • ProSim 8 Carrying Case (4034597)
    • ProSim NIBP MANDREL SET (4308086)
    • AC/DC Power Supply
    • AC Power cord Brazil
    • Ansur demo CD-ROM
    ProSim SPOT
    (3985658)
    ProSim SpO2 Test Module
    ProSim RAINBOW
    (4034609)
    ProSim RAINBOW Test CABLE

    Cables (IBP)
    3010-0368CM-3 BP cable (Nicolet round – 12 meter) (part# 2199682)
    GE Marquette Medical Corametrics (115, 116, 142, 145, 556)
    3010-0357GE Marquette Medical (Dash, Eagle, Solar, Tram and MacLab) MQ-3 BP cable (rectangular -11 M) (part# 2199627)
    3010-0370HP-8 BP cable (part# 2199694)
    Hewlett Packard/Philips (8040A, M1350A) HP-8 (intrauterine pressure only – 12M 40µV)
    MINIDIN/DIN IBPMini DIN to DIN IBP Adapter (part# 3408564)
    •Nihon Kohden
    Used to adapt PS420 or MPS450 blood pressure cables for use with ProSim series simulators.

    ProSim 8 Cable Kits
    3984993
    PS8 DRAGER CABLES
    Drager Infinity Cable Set (part# 3984993) for ProSim 8 includes:
    •SM-1 BP Cable (part# 2198925, two)
    •UT-4,Low profile 1/4 inch phone plug
    •YSI 400 Series Compatible 2 conductor (part# 2523334)
    3984968
    PS8 GE CABLES
    GE Marquette Eagle/Dash/Solar Cable Set (part# 3984968) for ProSim 8 includes:
    •MQ-3 BP Cable (part# 2199627, two)
    •UT-4,Low profile 1/4 inch phone plug
    •YSI 400 Series Compatible 2 conductor (part# 2523334)
    •UT-2 Tamp Cable 700 Series YSI (part# 2199019)
    •PROSIM8-4402GECO Din cardiac Output Marq Eagle (part# 4022300)
    3984922
    PS8 HP/PHI CABLES
    HP/Philips Intellivue Cable Set (part# 3984922) for ProSim 8 includes:
    •HP-3 BP Cable (part# 2198902, two)
    •HPT-2 Tamp/C.O. Injct Cable Assembly (part# 2199257)
    •COA-1 Cable Assembly (part# 2199240)
    •UT-4,Low profile 1/4 inch phone plug
    •YSI 400 Series Compatible 2 conductor (part# 2523334)
    3985009
    PS8 NK CABLES
    Nihon Kohden Cable Set (part# 3985009) for ProSim 8 includes:
    •Nihon Kohden-NK-1, BP Cable (5M) (part# 2462263, two)
    •DIN to minDin adapter (part# 3408564)
    •UT-4, Low profile 1/4 inch phone plug
    •YSI 400 Series Compatible 2 conductor (part# 2523334)
    3984946
    PS8 SPACEL CABLES
    SpaceLabs Ultraview Cable Set (part# 3984946) for ProSim 8 includes:
    •TK-1 BP Cable (part# 2198879, two)
    •UT-4,Low profile 1/4 inch phone plug
    •YSI 400 Series Compatible 2 conductor (part# 2523334)
    •UT-2 Tamp Cable 700 Series YSI (part# 2199019)
    •CI-3 Cable Assembly (part# 2392199)
    3984979
    PS8 WA/PRO CABLES
    Welch Allyn/Propaq Cable Set (part# 3984979) for ProSim 8 includes:
    •TK-1 BP Cable (part# 2198879, two)
    •UT-4,Low profile 1/4 inch phone plug
    •YSI 400 Series Compatible 2 conductor (part# 2523334)
    •UT-2 Tamp Cable 700 Series YSI (part# 2199019)

    Adapters
    2719-0153FG4-pin female audio connector (part# 2392158)
    General purpose cardiac device output connector
    3010-0289FGCI-3 Cardiac output test module (part# 2392199)

    Batteries, Chargers, and Power Supply
    BPPS68ProSim 6/8 Battery Pack

    Cables
    4034393Data Transfer Cable for use with ProSim 4, ProSim 8, ESA615
    3010-03434100-60 MQ-3 cable (part# 2199549)
    GE Marquette Medical (Dash, Eagle, Solar, Tram, and MacLab) MQ-3 (rectangular – 11M)
    3010-0284COA-1 Cardiac output adapter (Hewlett Packard)
    1/4 Phone plug to 4 pin female audio connector
    (HPT-2 also required for cardiac output simulation on HP patient-monitoring systems) (part# 2199240)
    3010-01033010-0103 Datascope DS-1 BP cable (6F) (part# 2198887)
    3010-0104EM-1 BP cable for ProSim 2/3/4/8 (6ft) (6-pin) (part# 2198893)
    For use with the following:
    •GE Marquette Medical (PPG/E for M DR)
    •Witt Biomedical
    •Schiller PB Series
    3010-0307Fukuda Denshi (DS3300 series) FD-2 BP cable (12M) (12-pin orange) (part# 2199387)
    3010-0285HPT-2 temperature adapter (2 pin) (part# 2199257)
    (Hewlett Packard/Philips)(used with UT-1 for HP monitors)
    (COA-1 also required for cardiac output simulation on HP patient-monitoring systems)
    3010-0110HP-3 BP cable (12M 5µV) 120 in, Mini-DIN to 12 pin HP/PH (part# 2198902)
    For use with the following:
    •Hewlett Packard/Philips (78-300, 78-500, 78-800, Merlin/Viridia/Omnicare)
    (HP/Philips M1006B iBP module has a sensitivity of 5 uv/V/mmHg only. Select the HP-3 cable for this application)
    3010-0111HP-4 BP cable (12M 40µV) for ProSim 8, MedSim 300B, 217A (part# 2198916)
    connecting to Hewlett Packard/Philips (78-300, 78-500, 78-800, Merlin/Viridia/Omnicare)
    3010-0115Mennen Medical (Horizon series) MM-1 (6M) (6-pin) (part# 2198933)
    3010-0122GE Marquette Medical (7000 and TRAM-AR series only) MQ-2 (8M round) (7-pin) (part# 2198978)
    3010-0116PC-1 BP cable (6M) (6-pin) (part# 2198940)
    For use with the following:
    •Physio Control (VSM series)
    •Medical Data Electronics (Escort series)
    ProSim8-4402GECOProSim8-4402GECO Cardiac Output Switch for GE (Marquette Eagle) (part# 4022300) for use with ProSim 2/3 and ProSim 8
    3010-0584DX-1 BP cable (10F) (10-pin) (part# 2200955)
    Used with the following:
    •Puritan Bennett PB 240
    •Datex (AS/3, CS/3, Compact, Cardio Cap II, Critical Care, Light)
    3010-0248QM-1 (6M) Blood pressure cable (10-pin orange) (part# 2198925)
    for use with the following:
    •Quinton (Q Cath series)
    •Drager/Siemens
    (Some of Drager/Siemens monitors use the standard 10 pin Siemen connector, but newer models require Drager's IBP Adapter 10 to 7 pin. Drager PN: 3368383)
    3010-0366SM-3 BP cable (15M) (part# 2199666)
    •Siemens (Micor/Mingo)
    3010-0114SM-1 BP cable (10M) for Pro Sim 2/3, 4 or ProSim 8 (10-pin) (part# 2198925)
    •Drager/Siemens (SIRECUST series)
    (Use the SM-1 and the Siemens Medical Transducer Adapter (3368-383-E530U) to run a single invasive BP channel on the Siemens Medical SC6000 and SC9000 series monitors)
    3010-00763010-0076 TK-1 (6M) BP Cable (part# 2198879) applies to:
    • BCI International
    • Criticare Systems Inc. (1100)
    • Critikon (Dinamap Plus)
    • Invivo Research
    • Ivy Biomedical (400 and 700 series)
    • North American Draqer (Vitalert 2000)
    • Protocol System (Propaq series) Zoll Defib
    • SpaceLabs (1050, 1700, PCMS series)
    (SpaceLabs adapters 700-0028-00 and 0120-0551-00 with TK1 used when testing the new UltraView Command Module)
    3010-0048FGUniversal unterminated UU-1 BP cable (5-Pin DIN one end only) (part# 2392173)
    3010-0193UT-2 Standard 1/4" phone plug (compatible with YSI 700 series – 3 conductor) temperature cable (part# 2199019)
    Use with the following:
    •Spacelabs Ultraview (use adapter# 700-0031-00)
    •GE Marquette (use adapter# 402015-005 or 402015-004)
    •Welch Allyn/Propaq
    UT-4UT-4 Low profile, ¼ in phone plug, YSI 400 series compatible, two conductor (part# 2523334 or 3015280)
    •HP/Philips
    •SpaceLabs Ultraview (adapter# 700-0031-00)
    •GE Marquette Eagle/Dash/Solar (adapter# 402015-005 or 402015-004)
    •Drager Infinity
    •Welch Allyn/Propaq
    •Nihon Kohden (requires Nihon Kohden Smart Temperature Cable K961 for the latest series TR-6000)

    Miscellaneous Accessories
    Ansur PROSIM 6/8Ansur ProSim Plug-In Software (part# 4034627)
    PROSIM 6/8 CASEProSim CARRYING CASE (part# 4034597)
    ProSim ChamberNIBP Test 500ML Rigid Chamber (part# 4034611)
    MANDREL-5001ProSim NIBP Mandrel Set (part# 4308086)
    90-183Wireless USB Dongle (part# 3341333)
    ZigBee

    Service Packages
    Care PlansRequest a quote
    Biomedical Equipment  |  Patient Simulator  |  Safety Analyzer  |  Dosimeter  |  Radiation Detector
    Search Fluke Biomedical
    ResourcesProducts
    Find Resources
    Related Products
    View related products
     
     ProSim 8 Accessory Kit »
     
     ProSim 4 Vital Signs Patient Simulator »
     
     PS410 Arrhythmia Patient Simulator »
    ProSim patient simulators for spo2, cardiac and vital sign device analysis
    ProSim 8 simulator training
    Patient simulator selection guide
    Fluke Biomedical ProSim Cable Selection Guide

        Follow us                      Home   |   Site Map   |   Fluke Corporation   |   Terms and Conditions   |   Privacy Statement   |   Disclaimer© 1995 - 2014 Fluke Corporation